DOI 10.1161/circinterventions.111.965111
Impact of frequency-domain optical coherence tomography guidance for optimal coronary stent implantation in comparison with intravascular ultrasound guidance.
Аннотация
BACKGROUND
Frequency-domain optical coherence tomography (FD-OCT) is a novel, high resolution intravascular imaging modality. Intravascular ultrasound (IVUS) is a widely used conventional imaging modality for achieving optimal stent deployment. The aim of this study was to evaluate the impact of FD-OCT guidance for coronary stent implantation compared with IVUS guidance.
METHODS AND
RESULTS
A total of 70 patients with de novo coronary artery lesions and either unstable or stable angina pectoris were enrolled in this randomized study (optical coherence tomography [OCT] group: n=35, IVUS group: n=35). In the OCT group, stent implantation was performed under FD-OCT guidance alone and final stent expansion was evaluated by IVUS. In the IVUS group, conventional IVUS guidance was used and final stent apposition was evaluated by FD-OCT. There were no significant differences regarding the procedural, fluoroscopy time, and contrast volume. Although device and clinical success rates also were similar, the visibility of vessel border was significantly lower in the OCT group (P<0.05). Minimum and mean stent area and focal and diffuse stent expansion were smaller (6.1±2.2 mm versus 7.1±2.1 mm, 7.5±2.5 versus 8.7±2.4 mm, 64.7±13.7% versus 80.3±13.4%, 84.2±15.8% versus 98.8±16.5%, P<0.05, respectively), and the frequency of significant residual reference segment stenosis at the proximal edge was higher in the OCT group (P<0.05). Incomplete apposed struts in both groups were similar (P=0.34).
CONCLUSIONS
FD-OCT guidance for stent implantation was associated with smaller stent expansion and more frequent significant residual reference segment stenosis compared with conventional IVUS guidance.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.