EMG pattern recognition compared to foot control of the DEKA Arm.

Клиническое исследованиеСравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Resnik Linda J, Acluche Frantzy, Borgia Matthew, Cancio Jill, Latlief Gail, Phillips Samuel, Sasson Nicole

Аннотация

INTRODUCTION: EMG pattern recognition control (EMG-PR) is a promising option for control of upper limb prostheses with multiple degrees of freedom (DOF). The purposes of this study were to 1) evaluate outcomes of EMG-PR and inertial measurement units (IMU) control of the DEKA Arm as compared to personal prosthesis; and 2) compare outcomes of EMG-PR to IMU control of DEKA Arm.

METHODS: This was a quasi-experimental, multi-site study with repeated measures that compared non-randomized groups using two types of controls: EMG-PR and IMUs. Subjects (N = 36) were transradial (TR) and transhumeral (TH) amputees. Outcomes were collected at Baseline (using personal prosthesis), and after in-laboratory training (Part A), and home use (Part B). Data was compared to personal prosthesis, stratified by amputation level and control type. Outcomes were also compared by control type.

RESULTS: The EMG-PR group had greater prosthesis use after Part A, but worse dexterity, lower satisfaction, and slower activity performance compared to Baseline; the IMU group had slower activity performance. After Part B, the EMG-PR group had less perceived activity difficulty; the IMU group had improved activity performance, improved disability and activity difficulty, but slower performance. No differences were observed for TH group by control type in Part A or B. The TR group using EMG-PR had worse dexterity (Parts A & B), and activity performance (Part A) as compared to IMU users.

DISCUSSION/CONCLUSION: Findings suggest that for the TR group that IMUs are a more effective control method for the DEKA Arm as compared to the EMG-PR prototypes employed in this study. Further research is needed to refine the EMG-PR systems for multi-DOF devices. Future studies should include a larger sample of TH amputees.

TRIAL REGISTRATION: ClinicalTrials.gov NCT01551420.