BACKGROUND

Colonic stents are indicated for preoperative and palliative decompression of large bowel obstruction. We aim to investigate post Food and Drug Administration (FDA) approval outcomes associated with colonic stents.

MATERIALS AND

METHODS

We analyzed postmarketing surveillance data from the FDA Manufacturer and User Facility Device Experience (MAUDE) database from January 2011 to December 2020.

RESULTS

During the study period, 691 device issues and 742 patient complications were identified. The number of adverse events reported to the FDA mildly increased from 6.40% in 2011 to 7.4% in 2020. Most device complications were due to a positioning problem (n=201, 29%), followed by device break (n=60, 8.6%), delamination (n=41, 6.3%), and migration or expulsion device (n=38, 5.55%). A number of reports described an unclassified adverse event without specifying device or operator problem (n=141, 20.3%). The most reported patient adverse events were perforation (n=150, 20.1%), death (n=47, 6.3%), abdominal pain/discomfort (n=27, 3.6%), and peritonitis (n=24, 3.2%).

CONCLUSION

Findings from the MAUDE database highlight rare but important patient and device complications which endoscopists and referring providers should be aware of before consideration for colonic stent placement.

Для цитирования: Ramai Daryl, DeLuca Matthew, Facciorusso Antonio, Ofosu Andrew, Girotra Mohit, Tabibian James H, Rodrigues-Pinto Eduardo, Adler Douglas G, Barakat Monique T. Analysis of Reported Adverse Events With Colonic Stents: An FDA MAUDE Database Study.. Journal of clinical gastroenterology. 2021. DOI: 10.1097/mcg.0000000000001632.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.