Dose Distribution Degradation of Carbon-ion Radiotherapy Caused by Tumor Cell-specific Relative Biological Effectiveness of Osteosarcoma: A Simulation Study Using In Vitro Experimental Results.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Shiba Shintaro, Kawashima Motohiro, Okamoto Masahiko, Ohno Tatsuya
Аннотация
BACKGROUND/AIM: Dose distributions of carbon-ion radiotherapy (C-ion RT) have been created with the relative biological effectiveness (RBE) of human salivary gland cells (HSG). However, no dose distributions have been created using various tumor cell-specific RBE values. Hence, we conducted in vitro experiments to determine the RBE of human osteosarcoma cells (U2OS) and used this RBE value (RBEU2OS) to calculate the dose distribution for C-ion RT.
MATERIALS AND METHODS: To obtain RBE values for various linear energy transfer (LET) levels, we exposed U2OS cells to different doses of X-rays and varying doses and LET levels of C-ion beams (13, 30, 50, and 70 keV/μm). Subsequently, we converted the RBE of HSG (RBEHSG) to RBEU2OS in the treatment planning system and reconstructed the dose distribution for a typical osteosarcoma case. We performed a dose-volume histogram (DVH) analysis, evaluating the percentage of the minimum dose that covered 98%, 50%, and 2% (D98%, D50%, and D2%, respectively), as well as the homogeneity index [HI; calculated as (D2%-D98%)/D50%].
RESULTS: The RBEU2OS values for C-ion beams with LET of 13, 30, 50, and 70 keV/μm were 1.77, 2.25, 2.72, and 4.50, respectively. When comparing DVH parameters with the planning target volume, we observed the following values: D98%, D50%, D2%, and HI for RBEHSG were 64.1, 70.1, 72.4 Gy (RBE), and 0.12, respectively. For RBEU2OS, these values were 86.2, 95.0, 107.9 Gy (RBE), and 0.23, respectively.
CONCLUSION: We utilized RBEU2OS to calculate the dose distribution of carbon ion radiotherapy, revealing potential degradation in dose distribution and particularly worsening of the HI.
MATERIALS AND METHODS: To obtain RBE values for various linear energy transfer (LET) levels, we exposed U2OS cells to different doses of X-rays and varying doses and LET levels of C-ion beams (13, 30, 50, and 70 keV/μm). Subsequently, we converted the RBE of HSG (RBEHSG) to RBEU2OS in the treatment planning system and reconstructed the dose distribution for a typical osteosarcoma case. We performed a dose-volume histogram (DVH) analysis, evaluating the percentage of the minimum dose that covered 98%, 50%, and 2% (D98%, D50%, and D2%, respectively), as well as the homogeneity index [HI; calculated as (D2%-D98%)/D50%].
RESULTS: The RBEU2OS values for C-ion beams with LET of 13, 30, 50, and 70 keV/μm were 1.77, 2.25, 2.72, and 4.50, respectively. When comparing DVH parameters with the planning target volume, we observed the following values: D98%, D50%, D2%, and HI for RBEHSG were 64.1, 70.1, 72.4 Gy (RBE), and 0.12, respectively. For RBEU2OS, these values were 86.2, 95.0, 107.9 Gy (RBE), and 0.23, respectively.
CONCLUSION: We utilized RBEU2OS to calculate the dose distribution of carbon ion radiotherapy, revealing potential degradation in dose distribution and particularly worsening of the HI.