m+Medelevator.RU
← Поиск публикаций
Tomography (Ann Arbor, Mich.) · 2023

Impact of Tracer Dose Reduction in [18 F]-Labelled Fluorodeoxyglucose-Positron Emission Tomography ([18 F]-FDG)-PET) on Texture Features and Histogram Indices: A Study in Homogeneous Tissues of Phantom and Patient.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Vogel Jonas, Seith Ferdinand, Estler Arne, Nikolaou Konstantin, Schmidt Holger, la Fougère Christian, Küstner Thomas

Аннотация

BACKGROUND: Histogram indices (HIs) and texture features (TFs) are considered to play an important role in future oncologic PET-imaging and it is unknown how these indices are affected by changes of tracer doses. A randomized undersampling of PET list mode data enables a simulation of tracer dose reduction. We performed a phantom study to compare HIs/TFs of simulated and measured tracer dose reductions and evaluated changes of HIs/TFs in the liver of patients with PETs from simulated reduced tracer doses. Overall, 42 HIs/TFs were evaluated in a NEMA phantom at measured and simulated doses (stepwise reduction of [18 F] from 100% to 25% of the measured dose). [18 F]-FDG-PET datasets of 15 patients were simulated from 3.0 down to 0.5 MBq/kgBW in intervals of 0.25 MBq/kgBW. HIs/TFs were calculated from two VOIs placed in physiological tissue of the right and left liver lobe and linear correlations and coefficients of variation analysis were performed.

RESULTS: All 42 TFs did not differ significantly in measured and simulated doses (p > 0.05). Also, 40 TFs showed the same behaviour over dose reduction regarding differences in the same group (measured or simulated), and for 26 TFs a linear behaviour over dose reduction for measured and simulated doses could be validated. Out of these, 13 TFs could be identified, which showed a linear change in TF value in both the NEMA phantom and patient data and therefore should maintain the same informative value when transferred in a dose reduction setting. Out of this Homogeneity 2, Entropy and Zone size non-uniformity are of special interest because they have been described as preferentially considerable for tumour heterogeneity characterization.

CONCLUSIONS: We could show that there was no significant difference of measured and simulated HIs/TFs in the phantom study and most TFs reveal a linear behaviour over dose reduction, when tested in homogeneous tissue. This indicates that texture analysis in PET might be robust to dose modulations.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты