Echocardiographic description of the CarboMedics bileaflet prosthetic heart valve.

Chambers J, Cross J, Deverall P, Sowton E

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to describe the echocardiographic appearance of the normal CarboMedics prosthesis in the aortic and mitral positions.

BACKGROUND

Echocardiography is the standard method of assessing prosthetic valves. However, new valve designs may still be marketed without an accompanying echocardiographic description. The CarboMedics prosthesis is in widespread use, but few noninvasive hemodynamic data have been published.

METHODS

Echocardiography was performed in 147 patients with a total of 96 normally functioning CarboMedics prostheses in the aortic position and 75 in the mitral position; in 24 patients, valves were implanted in both positions. The following variables were measured: peak and mean transvalvular velocities, peak and mean instantaneous gradient estimated from the modified Bernoulli equation, aortic acceleration slope, pressure half-time, transvalvular flow and effective orifice area using the continuity equation. Patterns of regurgitation were observed by transthoracic study in all valves and by transesophageal study in selected mitral valve prostheses.

RESULTS

For the aortic valve prostheses, estimated mean gradient ranged between 6 and 19 mm Hg. Effective area differed markedly among the anulus diameters (p < 0.001), with a mean value of 1 cm2 for the 19-mm valve and 2.6 cm2 for the 29-mm valve. For the mitral valve prostheses, mean gradient ranged from 3 to 7 mm Hg. There were a total of four washing leaks, one on either side of each pivotal point, and these lasted throughout systole or diastole. One jet was commonly more prominent than the other three.

CONCLUSIONS

The CarboMedics prosthesis offered relatively little resistance to forward flow except at small anulus diameters. The washing jets were prominent and would be easy to misdiagnose as a sign of paraprosthetic regurgitation.

Для цитирования: Chambers J, Cross J, Deverall P, Sowton E. Echocardiographic description of the CarboMedics bileaflet prosthetic heart valve.. Journal of the American College of Cardiology. 1993. DOI: 10.1016/0735-1097(93)90681-p.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.