[Ball variance and fracture of a Smeloff-Cutter prosthesis 24 years after aortic valve replacement].

Hust M H, Klinkmüller A, Keim M, Momper R, Nothwang G

Клинический случайАннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

This report documents a case of ball variance in a Smeloff-Cutter aortic prosthesis occurring 24 years after implantation. After episodes of embolic complications the patient died in acute shock. The silicone rubber ball showed several alterations including discoloration, grooving, cracking, swelling and subtotal fracture of the poppet. Terminal valvular malfunction was caused by complete thrombosis of the prosthesis. In most patients ball variance occurred during the first years after valve replacement; thus, the observed case is a very rare late complication of a ball-valve prosthesis.

Для цитирования: Hust M H, Klinkmüller A, Keim M, Momper R, Nothwang G. [Ball variance and fracture of a Smeloff-Cutter prosthesis 24 years after aortic valve replacement].. Zeitschrift fur Kardiologie. 1997. DOI: 10.1007/s003920050092.