MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1055/s-2007-1001421

Медицинская библиотека · Endoscopy · 1998

Is urinary bladder pressure a sensitive indicator of intra-abdominal pressure?

Yol S, Kartal A, Tavli S, Tatkan Y

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND STUDY AIMS: The accuracy of transurethral bladder catheter pressure in reflecting intra-abdominal pressure is well defined in experimental studies and case reports but not in controlled clinical trials.

PATIENTS AND

METHODS

We compared bladder pressure with insufflator pressure during laparoscopic cholecystectomy in 40 patients. Measurements were made at four pressure levels of the insufflator: 0, 5, 10 and 15 mmHg.

RESULTS

When the insufflator displayed 0, 5, 10, and 15 mmHg (0, 6.8, 13.6 and 20.4 cmH2O), and the mean bladder pressures measured 2.5 +/- 1.4, 7.3 +/- 1.5, 12.9 +/- 1.6 and 19.7 +/- 1.5 cmH2O, respectively. The two measurements correlated well with each other (r = 0.973, P < 0.0001).

CONCLUSION

We concluded that bladder pressure measured by transurethral catheter was equal to insufflator pressure during laparoscopy, and that this was a valid indicator of intra-abdominal pressure.

Для цитирования: Yol S, Kartal A, Tavli S, Tatkan Y. Is urinary bladder pressure a sensitive indicator of intra-abdominal pressure?. Endoscopy. 1998. DOI: 10.1055/s-2007-1001421.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.