DOI 10.1111/jce.13654

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2018

First clinical use of novel ablation catheter incorporating local impedance data.

Martin Claire A, Martin Ruairidh, Gajendragadkar Parag R, Maury Philippe, Takigawa Masateru, Cheniti Ghassen, Frontera Antonio, Kitamura Takeshi, Duchateau Josselin, Vlachos Konstantinos, Bourier Felix, Lam Anna, Lord Stephen, Murray Stephen, Shephard Ewen, Pambrun Thomas, Denis Arnaud, Derval Nicolas, Hocini Meleze, Haissaguerre Michel, Jais Pierre, Sacher Frederic

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

Successful catheter ablation is limited by both poor spatial resolution of abnormal local signals and inability to deliver an effective lesion due to poor tissue contact. We report first worldwide use of the Intellanav MiFi OI catheter (Boston Scientific), providing ultra-high density mapping and incorporating a "DirectSense" algorithm to measure local tissue impedance (LI).

METHODS AND

RESULTS

31 patients (65±6 years, 20 male) underwent ablation. LI from the catheter, generator impedance (GI) and maximum electrogram amplitude were recorded in the blood pool, and in regions from healthy to dense scar before, during and after ablation. The catheter demonstrated clear nearfield signal where standard bipolar recordings included farfield signal. LI was lower in dense scar than either healthy tissue or blood pool, and demonstrated an exponential relationship with maximum electrogram amplitude. Maximum LI drop on ablation linearly correlated with initial LI. The median LI drop for successful lesions, resulting in lack of local tissue capture, was 16.0Ω (12.1-19.8 Ω) for LV and 14.6 Ω (10.0-18.3 Ω) for LA, which was larger than for unsuccessful lesions (LV: 9.4 Ω [5.4-15.6 Ω] P = 0.001; LA: 6.8 Ω [4.7-13.0 Ω], P = 0.049). LI percentage drop was also significantly larger for successful than unsuccessful lesions (LV: 17.1 Ω [14.0-19.6 Ω] vs. 10.6 Ω (7.1-16.5 Ω) P = 0.002; LA: 14.2 Ω [10.8-19.5 Ω] vs. 7.5Ω [5.1-11.0 Ω], P = 0.005).

CONCLUSION

This novel catheter gives reproducible recordings of local impedance, which are dependent on scar level. Absolute LI drop, and also percentage drop, on ablation may give an indication of tissue contact and subsequent effective lesion formation.

Для цитирования: Martin Claire A, Martin Ruairidh, Gajendragadkar Parag R, Maury Philippe, Takigawa Masateru, Cheniti Ghassen, Frontera Antonio, Kitamura Takeshi, Duchateau Josselin, Vlachos Konstantinos, Bourier Felix, Lam Anna, Lord Stephen, Murray Stephen, Shephard Ewen, Pambrun Thomas, Denis Arnaud, Derval Nicolas, Hocini Meleze, Haissaguerre Michel, Jais Pierre, Sacher Frederic. First clinical use of novel ablation catheter incorporating local impedance data.. Journal of cardiovascular electrophysiology. 2018. DOI: 10.1111/jce.13654.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.