DOI 10.1016/j.jmir.2016.05.002

Медицинская библиотека · Journal of medical imaging and radiation sciences · 2016

Correct Patient Centering Increases Image Quality without Concomitant Increase of Radiation Dose during Adult Intracranial Computed Tomography.

Almohiy Hussain, Alasar Emadeldin Mohamed Mukhtar, Saade Charbel

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the impact of patient centering and radiation dose during intracranial computed tomography (ICT) on quantitative and qualitative image quality.

MATERIALS AND

METHODS

A total of 500 consecutive patients who underwent ICT were retrospectively reviewed using a 128-slice CT scanner (Definition AS+, Siemens, Germany). Patients were subjected in equal numbers to one of two positioning protocols: group A, poorly centered; and group B involved accurate centering before imaging. Gray-white matter (GWM) conspicuity, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) in each group were calculated. Qualitative image quality in terms of GWM differentiation, distinctness of posterior fossa contents, and overall diagnostic acceptability were evaluated by 2 neuroradiologists. The dose length product, CNR, SNR, and noise were measured between each group and data generated were compared using Mann-Whitney U nonparametric statistics. Visual grading characteristic and Kappa analyses were performed.

RESULTS

The mean noise index was significantly lower in group B (2.61 ± 0.29) compared with A (2.66 ± 0.21; P < .02). The mean attenuation of GWM, SNR, and CNR in the frontal lobe (A, 1:0.77, 0.84, 8.70 ± 1.36; and B, 1:0.65, 0.85, 15.32 ± 1.21; P < .02), occipital lobe (A, 1:1.10, 1.18, 10.79±2.11; and B, 1:0.94, 0.64, 14.41 ± 3.09; P < .04), and cerebellum (A, 1:0.79, 0.90, 12.56 ± 4.08; and B, 1:0.82, 0.87, 14.07 ± 2.28; P < .04) were significantly higher in group B compared with A, while the globus pallidus, caudate nucleus, and optic track in the basal ganglia demonstrated no difference in each group (P > .05). Mean dose length product demonstrated no significance between each group (A, 1312.03 ± 133.92; B, 1298.11 ± 130.61). The qualitative analyses demonstrated significant increases in visual grading characteristic for each reader (P < .02) and interobserver agreement was significantly increased in protocol B (k = 0.81) compared with A (k = 0.62).

CONCLUSIONS

Correct patient centering increases the CNR and SNR in both GWM in the left and right hemispheres of the brain during ICT.

Для цитирования: Almohiy Hussain, Alasar Emadeldin Mohamed Mukhtar, Saade Charbel. Correct Patient Centering Increases Image Quality without Concomitant Increase of Radiation Dose during Adult Intracranial Computed Tomography.. Journal of medical imaging and radiation sciences. 2016. DOI: 10.1016/j.jmir.2016.05.002.