DOI 10.1016/j.jcin.2012.11.006
Clinical outcome of patients with and without diabetes mellitus after percutaneous coronary intervention with the resolute zotarolimus-eluting stent: 2-year results from the prospectively pooled analysis of the international global RESOLUTE program.
Аннотация
OBJECTIVES
The aim of this study was to describe the process to obtain Food and Drug Administration (FDA) approval for the expanded indication for treatment with the Resolute zotarolimus-eluting stent (R-ZES) (Medtronic, Inc., Santa Rosa, California) in patients with coronary artery disease and diabetes.
BACKGROUND
The R-ZES is the first drug-eluting stent specifically indicated in the United States for percutaneous coronary intervention in patients with diabetes.
METHODS
We pooled patient-level data for 5,130 patients from the RESOLUTE Global Clinical Program. A performance goal prospectively determined in conjunction with the FDA was established as a rate of target vessel failure at 12 months of 14.5%. In addition to the FDA pre-specified cohort of less complex patients with diabetes (n = 878), we evaluated outcomes of the R-ZES in all 1,535 patients with diabetes compared with all 3,595 patients without diabetes at 2 years.
RESULTS
The 12-month rate of target vessel failure in the pre-specified diabetic cohort was 7.8% (upper 95% confidence interval: 9.51%), significantly lower than the performance goal of 14.5% (p < 0.001). After 2 years, the cumulative incidence of target lesion failure in patients with noninsulin-treated diabetes was comparable to that of patients without diabetes (8.0% vs. 7.1%). The higher risk insulin-treated population demonstrated a significantly higher target lesion failure rate (13.7%). In the whole population, including complex patients, rates of stent thrombosis were not significantly different between patients with and without diabetes (1.2% vs. 0.8%).
CONCLUSIONS
The R-ZES is safe and effective in patients with diabetes. Long-term clinical data of patients with noninsulin-treated diabetes are equivalent to patients without diabetes. Patients with insulin-treated diabetes remain a higher risk subset. (The Medtronic RESOLUTE Clinical Trial; NCT00248079; Randomized, Two-arm, Non-inferiority Study Comparing Endeavor-Resolute Stent With Abbot Xience-V Stent [RESOLUTE-AC]; NCT00617084; The Medtronic RESOLUTE US Clinical Trial (R-US); NCT00726453; RESOLUTE International Registry: Evaluation of the Resolute Zotarolimus-Eluting Stent System in a 'Real-World' Patient Population [R-Int]; NCT00752128; RESOLUTE Japan-The Clinical Evaluation of the MDT-4107 Drug-Eluting Coronary Stent [RJ]; NCT00927940).