MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.24907

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2013

Neointimal tissue component assessed by tissue characterization with 40 MHz intravascular ultrasound imaging: comparison of drug-eluting stents and bare-metal stents.

Tsujita Kenichi, Takaoka Naoko, Kaikita Koichi, Hokimoto Seiji, Horio Eiji, Sato Koji, Mizobe Michio, Nakayama Naoki, Kojima Sunao, Tayama Shinji, Sugiyama Seigo, Nakamura Sunao, Ogawa Hisao

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The present study used iMap IVUS system to compare neointimal tissue components between DES and bare-metal stents (BMSs).

BACKGROUND

Drug-eluting stents (DESs) can cause impaired arterial healing, which constitutes the most important pathological substrate underlying late DES thrombosis. Intravascular ultrasound (IVUS)-based tissue characterization allows for the in vivo identification of neointimal tissue components.

METHODS AND

RESULTS

Follow-up IVUS data after coronary stenting (9.8 ± 9.4 months from index procedures) was obtained from consecutive 61 lesions (34 in DES, 27 in BMS). The iMap tissue components (fibrotic, lipidic, necrotic, and calcified) were measured in every recorded frame and expressed as percentages of mean neointimal cross-sectional area for the stented segment. Patients' characteristics were comparable between DES and BMS. When compared with BMSs, smaller (2.9 ± 0.4 mm vs. 3.2 ± 0.4 mm, P = 0.004) and longer (34 ± 18 mm vs. 26 ± 14 mm, P = 0.03) DESs were implanted. When compared with BMS group, minimum lumen area at follow-up was significantly greater in DES group (3.9 ± 1.8 mm(2) vs. 3.1 ± 1.5 mm(2) , P < 0.04), mainly attributable to suppression of neointimal hyperplasia (1.7 ± 0.8 mm(2) vs. 3.1 ± 1.5 mm(2) , P < 0.0001). The iMap analyses showed that neointima after DES placement was composed of smaller fibrotic component (67 ± 8% vs. 78 ± 7%, P < 0.0001), larger necrotic (14 ± 4% vs. 9 ± 3%, P < 0.0001) and calcified (15 ± 6% vs. 7 ± 4%, P < 0.0001) components compared with BMS. Logistic regression analysis showed that only intra-DES neointima was a significant predictor of necrotic neointima at follow-up.

CONCLUSIONS

DES implantation would be associated with iMap-derived necrotic and less-fibrotic neointimal formation. In vivo iMap evaluation of neointimal tissue may provide useful information in detecting impaired healing after stenting.

Для цитирования: Tsujita Kenichi, Takaoka Naoko, Kaikita Koichi, Hokimoto Seiji, Horio Eiji, Sato Koji, Mizobe Michio, Nakayama Naoki, Kojima Sunao, Tayama Shinji, Sugiyama Seigo, Nakamura Sunao, Ogawa Hisao. Neointimal tissue component assessed by tissue characterization with 40 MHz intravascular ultrasound imaging: comparison of drug-eluting stents and bare-metal stents.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2013. DOI: 10.1002/ccd.24907.