Slot fidelity and ligation-dependent tribology in personalized brackets made by casting versus selective laser melting: an in vitro study.

Nguyen Viet Anh, Tran Minh Ngoc, Pham Ngo The Minh, Ha Thi Bich Ngoc, Hoang Viet, Vuong Thi Quynh Trang

СтатьяПеревод пока не подготовлен

Аннотация

Reliable chairside adoption of digital orthodontics depends on micrometre-scale slot fidelity and stable bracket-wire tribology. A single computer-aided design for a personalized second-premolar bracket was manufactured by lost-wax casting and by selective laser melting (SLM) (n = 36 per method). Slot height and inter-wall angle were measured on both the support-facing and non-support surfaces. Static and dynamic friction were evaluated using stainless-steel rectangular wires ligated either with tightly twisted stainless-steel ties or with elastomeric modules. Mean slot height was 480.88 ± 73.90 µm for casting and 421.47 ± 32.03 µm for SLM, against a nominal 480 µm. Overall height error did not differ between methods (P = 0.673), whereas angle er-ror was greater for SLM (17.76 ± 11.29°) than for casting (9.56 ± 8.88°, P < 0.001). The support-facing wall consistently showed reduced accuracy: in casting, height error in-creased on the support side (P = 0.001); in SLM, both height and angle errors increased on the support side (both P < 0.001). Across conditions, static friction exceeded dynamic friction (all P < 0.001). With steel ligation, friction was higher for casting than SLM (median static 6.00 N vs 4.35 N, P = 0.007; median dynamic 5.03 N vs 3.83 N, P = 0.011). With elastomeric ligation, the ranking reversed, with higher friction for SLM (median static 2.95 N vs 2.05 N, P < 0.001; median dynamic 2.54 N vs 2.03 N, P = 0.003). In this standardized local in vitro model, findings suggest a surface-dependent reduction in accuracy at the support-facing wall and a fabrication-by-ligation interaction that may alter the frictional ranking between steel and elastomeric ligations. Cast brackets showed more consistent full-wire seating, whereas SLM brackets may benefit from calibrated design offsets and targeted finishing to limit undersized or tapered slots; validation beyond this setup is needed.

Для цитирования: Nguyen Viet Anh, Tran Minh Ngoc, Pham Ngo The Minh, Ha Thi Bich Ngoc, Hoang Viet, Vuong Thi Quynh Trang. Slot fidelity and ligation-dependent tribology in personalized brackets made by casting versus selective laser melting: an in vitro study.. Scientific reports. 2026. DOI: 10.1038/s41598-025-34810-w.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.