Long-term vascular healing in response to sirolimus- and paclitaxel-eluting stents: an optical coherence tomography study.

Räber Lorenz, Baumgartner Sandro, Garcia-Garcia Hector M, Kalesan Bindu, Justiz Jörn, Pilgrim Thomas, Moschovitis Aris, Khattab Ahmed A, Buellesfeld Lutz, Wenaweser Peter, Meier Bernhard, Serruys Patrick W, Jüni Peter, Windecker Stephan

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study sought to assess stent strut coverage, malapposition, protrusion, and coronary evaginations as markers of healing 5 years after implantation of sirolimus-eluting stents (SES) and paclitaxel-eluting stents (PES), by optical coherence tomography (OCT).

BACKGROUND

Early-generation drug-eluting stents have been shown to delay vascular healing.

METHODS

A total of 88 event-free patients with 1 randomly selected lesion were suitable for final OCT analysis 5 years after drug-eluting stent implantation. The analytical approach was based on a hierarchical Bayesian random-effects model.

RESULTS

OCT analysis was performed at 5 years in 41 SES lesions with 6,380 struts, and in 47 PES lesions with 6,782 struts. A total of 196 struts were uncovered in SES (1.5%) compared with 185 struts in PES lesions (1.0%, 95% credibility interval [CrI]: 0.5 to 1.6; p = 0.32). Malapposed struts were present in 1.2% of SES compared with 0.7% of PES struts (0.7%, 95% CrI: 0.03 to 1.6; p = 0.23). Protruding struts were more frequent among SES (n = 114; 0.8%) than PES lesions (n = 24; 0.1%, 95% CrI: 0.3 to 1.3; p < 0.01). Coronary evaginations were more common among SES- than PES-treated lesions (17 vs. 7 per 100 cross sections, p = 0.003). During extended clinical follow-up, 2 patients suffered from very late stent thrombosis showing a high degree of malapposition, protrusion, and coronary evaginations at the time of OCT investigation.

CONCLUSIONS

Early-generation drug-eluting stents show a similar degree of strut coverage and malapposition at 5 years of follow-up. Despite an overall low degree of uncovered and malapposed struts in event-free patients, some lesions show a clustering of these characteristics, indicating a heterogeneous healing response, which may be the source for very late adverse events.

Для цитирования: Räber Lorenz, Baumgartner Sandro, Garcia-Garcia Hector M, Kalesan Bindu, Justiz Jörn, Pilgrim Thomas, Moschovitis Aris, Khattab Ahmed A, Buellesfeld Lutz, Wenaweser Peter, Meier Bernhard, Serruys Patrick W, Jüni Peter, Windecker Stephan. Long-term vascular healing in response to sirolimus- and paclitaxel-eluting stents: an optical coherence tomography study.. JACC. Cardiovascular interventions. 2012. DOI: 10.1016/j.jcin.2012.05.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.