Pump Versus Syringe: Aspiration Thrombectomy Direct Pressure Comparisons in a Comprehensive Benchtop 3-Dimensional-Printed Circle of Willis Model.

СтатьяСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Berns Holly, Han Kimberly, Ramirez-Velandia Felipe, Lewis Kailey L, Robertson Sophia E, Clark Wyatt E, Wells Jesse C, Alnajrani Mohammed K, Rapoport Christopher, Schwartz Steven, Hakes Kana'i, Granstein Justin H, Ogilvy Christopher S, Becker Timothy A, Young Michael

Аннотация

BACKGROUND AND OBJECTIVES: Experiments have shown that 60-cc syringes generate stronger vacuum forces than mechanical pumps. However, few studies have measured catheter tip suction pressure during thrombus engagement. Therefore, this study aimed to compare aspiration forces and pressure drops, generated by mechanical pumps and 60-cc syringes connected to various catheters when aspirating synthetic clots in a 3-dimensional-printed model.

METHODS: The system includes a programmable pulsatile pump, a 3-dimensional-printed tissue-matched circle of Willis model, radiopaque synthetic blood clots positioned in the left internal carotid artery terminus, and pressure recording at 500 ± 10% mL/min and 120/80 ± 10 mm Hg. Each catheter underwent 3 soft and hard clot removal trials using a 60-cc syringe and an aspiration pump. Aspiration source pressure was recorded before, during, and after aspiration. The evaluated variables included engagement pressure change (∆P engagement ), aspiration pressure drops (∆P aspiration ), and pressure efficiency (P E ), defined as the percentage of peak pressure maintained during aspiration.

RESULTS: Aspirating soft clots with syringes showed low engagement pressure drops (∆P engagement ) (0-14 mm Hg), high P E at 93%-99% of P p, and minimal aspiration pressure drops (∆P aspiration ) (6F catheters: 4-8 mm Hg, 8F catheters: 19 mm Hg). Conversely, using a pump generated moderate ∆P engagement (6F: 16-51 mm Hg, 8F: 122 mm Hg), lower P E (69%-93% of P p ), and greater ∆P aspiration (6F: 143-151 mm Hg, 8F: 270 mm Hg). For hard clots, syringes exhibited minimal ∆P engagement and ∆P aspiration (4-13 mm Hg and 3-12 mm Hg, respectively), whereas pumps showed moderate ∆P engagement (2-34 mm Hg) and large ∆P aspiration (6F: 82-162 mm Hg, 8F: 280 mm Hg).

CONCLUSION: Syringe aspiration generates higher vacuum pressures and smaller pressure drops than aspiration pumps, likely because of vacuum tubing resistance in aspiration pumps. This effect is particularly evident with hard clots and larger inner diameter catheters.