DOI 10.1016/j.jvssci.2025.100287

Медицинская библиотека · JVS-vascular science · 2025

Quantifying the effect of IMPEDE-FX packing rate and volume on pressure-normalized principal wall strain in an idealized 3D-printed aneurysm model.

Kedwai Baqir, Geiger Joshua, Najjar Sam, Kruger Joel, Richards Michael, Yeh Chung, Dennehy Mary, Stoner Michael, Mix Doran

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

This study aimed to quantify the nonbiologic effects of Shape Memory IMPEDE-FX embolization plug deployment rate and packing volume on pressure-normalized wall strain ( ε ρ + ¯ /PP) of an idealized 3D-printed abdominal aortic aneurysm model.

METHODS

An endograft was deployed into an abdominal aortic aneurysm model and connected to an industry-validated hemodynamic simulator. Plugs were deployed into the excluded sac to packing volumes of 100%, 200%, 300%, and 400% under two conditions: (1) sequential and (2) immediate deployment. Axial ultrasound images were taken for each packing volume. Frame-to-frame displacements of the aneurysm wall were measured with ultrasound elastography over one cardiac cycle and normalized to the circuit's pulse pressure to calculate the mean principal strain ( ε ρ + ¯ /PP).

RESULTS

In the 100% packing condition, ε ρ + ¯ /PP was +113% above baseline at 15 minutes. After sequential deployment to 400%, the ε ρ + ¯ /PP trended down to +43% above baseline. Immediate packing was associated with a greater ε ρ + ¯ /PP reduction than sequential packing. When packed immediately to 400%, the ε ρ + ¯ /PP was -6.7% below baseline.

CONCLUSIONS

These modeling data suggest that an immediate deployment strategy and higher plug packing volumes are associated with lower ε ρ + ¯ /PP, which has been associated with decreased sac growth rates.

CLINICAL RELEVANCE: The present findings suggest that rapid, high-volume filling of IMPEDE-FX embolization plugs results in a reduction in wall ε ρ + ¯ /PP, independent of thrombus formation. Fully expanded embolization plugs in aggregate limit pulsatile aortic sac displacement likely contribute to a greater reduction in overall wall strain compared with low packing volumes. These findings may inform clinical application for this device, supporting a rapid and high-volume deployment strategy for greater reduction in ε ρ + ¯ /PP status post endovascular aneurysm repair.

Для цитирования: Kedwai Baqir, Geiger Joshua, Najjar Sam, Kruger Joel, Richards Michael, Yeh Chung, Dennehy Mary, Stoner Michael, Mix Doran. Quantifying the effect of IMPEDE-FX packing rate and volume on pressure-normalized principal wall strain in an idealized 3D-printed aneurysm model.. JVS-vascular science. 2025. DOI: 10.1016/j.jvssci.2025.100287.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.