MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s13550-025-01351-2

Медицинская библиотека · EJNMMI research · 2025

Renal dynamic scintigraphy using 3D-ring CZT cameras: comparison with NaI(Tl).

Metrard Gilles, Rouzic Gilles Le, Chapelle Benoit, Darsin-Bettinger Diane, Dekeyser Manon, Bailly Matthieu

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Isotopic renography is pivotal for assessing upper urinary tract obstruction and provides renal functional parameters. New 3D-CZT cameras offer enhanced sensitivity and innovative full SPECT protocols that need to be validated on phantoms. Current renal phantoms are often complex and suitable for 2D imaging. Using a new printed 3D renal phantom, we compared planar and full SPECT dynamic imaging.

METHODS

A 3D-printed renal phantom mimicking pediatric/adult kidneys was designed. Each kidney subunit, connected to a bladder-bag, was programmably infused with 7.5 MBq of [99mTc]-Pertechnetate to simulate normal nephrograms. An asymmetry of renal activity was simulated using different activity ratios. Each acquisition was conducted on both conventional NaI(Tl) and 3D-CZT cameras.

RESULTS

The 3D printed phantom enabled the acquisition of consistent and reproducible nephrograms. Tmax, T1/2, and A20 - min/max ratios showed no statistically significant differences between planar and SPECT. Simulated asymmetric renal function demonstrated a strong linear correlation between activity and AUC with r = 0.9455 and 0.9471 (p < 0.0001), respectively for conventional and 3D cameras.

CONCLUSION

This innovative 3D phantom allowed the acquisition of reproducible nephrograms with parameters comparable to those of clinical examinations. At low activity, 3D-CZT camera acquisitions provided equivalent values to conventional cameras, supporting their use for qualitative imaging and follow-up.

Для цитирования: Metrard Gilles, Rouzic Gilles Le, Chapelle Benoit, Darsin-Bettinger Diane, Dekeyser Manon, Bailly Matthieu. Renal dynamic scintigraphy using 3D-ring CZT cameras: comparison with NaI(Tl).. EJNMMI research. 2025. DOI: 10.1186/s13550-025-01351-2.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.