DOI 10.1111/1754-9485.12546

Медицинская библиотека · Journal of medical imaging and radiation oncology · 2016

Accurate tissue characterization in low-dose CT imaging with pure iterative reconstruction.

Murphy Kevin P, McLaughlin Patrick D, Twomey Maria, Chan Vincent E, Moloney Fiachra, Fung Adrian J, Chan Faimee E, Kao Tafline, O'Neill Siobhan B, Watson Benjamin, O'Connor Owen J, Maher Michael M

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

We assess the ability of low-dose hybrid iterative reconstruction (IR) and 'pure' model-based IR (MBIR) images to maintain accurate Hounsfield unit (HU)-determined tissue characterization.

METHODS

Standard-protocol (SP) and low-dose modified-protocol (MP) CTs were contemporaneously acquired in 34 Crohn's disease patients referred for CT. SP image reconstruction was via the manufacturer's recommendations (60% FBP, filtered back projection; 40% ASiR, Adaptive Statistical iterative Reconstruction; SP-ASiR40). MP data sets underwent four reconstructions (100% FBP; 40% ASiR; 70% ASiR; MBIR). Three observers measured tissue volumes using HU thresholds for fat, soft tissue and bone/contrast on each data set. Analysis was via SPSS.

RESULTS

Inter-observer agreement was strong for 1530 datapoints (rs > 0.9). MP-MBIR tissue volume measurement was superior to other MP reconstructions and closely correlated with the reference SP-ASiR40 images for all tissue types. MP-MBIR superiority was most marked for fat volume calculation - close SP-ASiR40 and MP-MBIR Bland-Altman plot correlation was seen with the lowest average difference (336 cm3 ) when compared with other MP reconstructions.

CONCLUSIONS

Hounsfield unit-determined tissue volume calculations from MP-MBIR images resulted in values comparable to SP-ASiR40 calculations and values that are superior to MP-ASiR images. Accuracy of estimation of volume of tissues (e.g. fat) using segmentation software on low-dose CT images appears optimal when reconstructed with pure IR.

Для цитирования: Murphy Kevin P, McLaughlin Patrick D, Twomey Maria, Chan Vincent E, Moloney Fiachra, Fung Adrian J, Chan Faimee E, Kao Tafline, O'Neill Siobhan B, Watson Benjamin, O'Connor Owen J, Maher Michael M. Accurate tissue characterization in low-dose CT imaging with pure iterative reconstruction.. Journal of medical imaging and radiation oncology. 2016. DOI: 10.1111/1754-9485.12546.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.