The Impella Recover 2.5 and TandemHeart ventricular assist devices are safe and associated with equivalent clinical outcomes in patients undergoing high-risk percutaneous coronary intervention.

Kovacic Jason C, Nguyen Huy T, Karajgikar Rucha, Sharma Samin K, Kini Annapoorna S

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To compare the practical use, safety, and clinical outcomes associated with the TandemHeart (TH) versus Impella Recover 2.5 (IR2.5) devices when used for circulatory support during high-risk percutaneous coronary intervention (PCI).

BACKGROUND

Small studies and registries suggest safety and efficacy for the TH and IR2.5 percutaneous-left ventricular assist devices (P-LVADs). However, these P-LVADs differ markedly in their insertion, operation, and manner of circulatory augmentation. To date, no study has compared these devices.

METHODS

We identified 68 patients (49 males, 19 females; age 71.1 ± 12.1 years) from our single-center database that underwent "high-risk" PCI with P-LVAD support from April 2005 to June 2010 (32 with TH, 36 with IR2.5). Relevant data were extracted for analysis.

RESULTS

Baseline demographics were similar, including low LVEF (overall mean 31.0 ± 13.7%) and elevated STS mortality risk score (4.2 ± 3.7%). Angiographic characteristics were also similar, with a mean of 2.4 ± 1.0 lesions treated per patient, and 29% undergoing left main PCI. PCI success rates were 99% in both groups, with similar in-hospital outcomes and a combined 7% major vascular access site complication rate. A single episode of left atrial perforation occurred during TH use. No patient required emergent CABG and no in-hospital deaths occurred. The 30-day MACE rate (death, myocardial infarction, target lesion revascularization) was 5.8%. There were no differences between the IR2.5 and TH groups with respect to short- or long-term clinical outcomes.

CONCLUSIONS

The IR2.5 and TH assist devices are safe, equally effective, and associated with acceptable short- and long-term clinical outcomes in patients undergoing "high-risk" PCI.

Для цитирования: Kovacic Jason C, Nguyen Huy T, Karajgikar Rucha, Sharma Samin K, Kini Annapoorna S. The Impella Recover 2.5 and TandemHeart ventricular assist devices are safe and associated with equivalent clinical outcomes in patients undergoing high-risk percutaneous coronary intervention.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2013. DOI: 10.1002/ccd.22929.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.