MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2010.07.008

Медицинская библиотека · JACC. Cardiovascular interventions · 2010

Improved late clinical safety with zotarolimus-eluting stents compared with paclitaxel-eluting stents in patients with de novo coronary lesions: 3-year follow-up from the ENDEAVOR IV (Randomized Comparison of Zotarolimus- and Paclitaxel-Eluting Stents in Patients With Coronary Artery Disease) trial.

Leon Martin B, Nikolsky Eugenia, Cutlip Donald E, Mauri Laura, Liberman Henry, Wilson Hadley, Patterson John, Moses Jeffrey, Kandzari David E, ENDEAVOR IV Investigators

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

The increased frequency of very late (>1 year) stent thrombosis (VLST) has raised concerns with regard to the safety of sirolimus-eluting stents and paclitaxel-eluting stents (PES).

BACKGROUND

Experimental and preliminary clinical findings with the zotarolimus-eluting stent (ZES) have suggested a favorable safety profile.

METHODS

The ENDEAVOR IV (Randomized Comparison of Zotarolimus- and Paclitaxel-Eluting Stents in Patients With Coronary Artery Disease) trial is a single-blind randomized ZES versus PES clinical trial in 1,548 patients with de novo native coronary lesions; the primary end point-9-month target vessel failure-was previously reported, annual clinical follow-up is planned for 5 years, and this report describes the 3-year outcomes.

RESULTS

The ZES compared with PES reduced target vessel failure (12.3% vs. 15.9%, hazard ratio [HR]: 0.76, 95% confidence interval [CI]: 0.58 to 1.00, p = 0.049), myocardial infarctions (MI) (2.1% vs. 4.9%, HR: 0.44, 95% CI: 0.25 to 0.80, p = 0.005), and cardiac death plus MI (3.6% vs. 7.1%, HR: 0.52, 95% CI 0.32 to 0.82, p = 0.004). Although the overall 3-year rate of Academic Research Consortium definite/probable stent thrombosis did not differ significantly (1.1% vs. 1.7%, HR: 0.67, 95% CI 0.28 to 1.64, p = 0.380), VLST (between 1 and 3 years) was significantly reduced in ZES patients (1 event vs. 11 events; 0.1% vs. 1.6%, HR: 0.09, 95% CI: 0.01 to 0.71, p = 0.004). Ischemia-driven target lesion revascularization at 3 years was similar with ZES versus PES (6.5% vs. 6.1%, HR: 1.10, 95% CI: 0.73 to 1.65, p = 0.662).

CONCLUSIONS

Three-year follow-up results from the ENDEAVOR IV trial indicate similar antirestenosis efficacy but improved clinical safety associated with ZES compared with PES, due to significantly fewer peri-procedural and remote MIs associated with fewer VLST events. (A Randomized, Controlled Trial of the Medtronic Endeavor Drug [ABT-578] Eluting Coronary Stent System Versus the Taxus Paclitaxel-Eluting Coronary Stent System in De Novo Native Coronary Artery Lesions; NCT00217269).

Для цитирования: Leon Martin B, Nikolsky Eugenia, Cutlip Donald E, Mauri Laura, Liberman Henry, Wilson Hadley, Patterson John, Moses Jeffrey, Kandzari David E, ENDEAVOR IV Investigators. Improved late clinical safety with zotarolimus-eluting stents compared with paclitaxel-eluting stents in patients with de novo coronary lesions: 3-year follow-up from the ENDEAVOR IV (Randomized Comparison of Zotarolimus- and Paclitaxel-Eluting Stents in Patients With Coronary Artery Disease) trial.. JACC. Cardiovascular interventions. 2010. DOI: 10.1016/j.jcin.2010.07.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.