A European multicenter study on the analytical performance of the VERIS HBV assay.
Аннотация
BACKGROUND
Hepatitis B viral load monitoring is an essential part of managing patients with chronic Hepatits B infection. Beckman Coulter has developed the VERIS HBV Assay for use on the fully automated Beckman Coulter DxN VERIS Molecular Diagnostics System.1
OBJECTIVES
To evaluate the analytical performance of the VERIS HBV Assay at multiple European virology laboratories.
STUDY
DESIGN
Precision, analytical sensitivity, negative sample performance, linearity and performance with major HBV genotypes/subtypes for the VERIS HBV Assay was evaluated.
RESULTS
Precision showed an SD of 0.15 log10 IU/mL or less for each level tested. Analytical sensitivity determined by probit analysis was between 6.8-8.0 IU/mL. Clinical specificity on 90 unique patient samples was 100.0%. Performance with 754 negative samples demonstrated 100.0% not detected results, and a carryover study showed no cross contamination. Linearity using clinical samples was shown from 1.23-8.23 log10 IU/mL and the assay detected and showed linearity with major HBV genotypes/subtypes.
CONCLUSIONS
The VERIS HBV Assay demonstrated comparable analytical performance to other currently marketed assays for HBV DNA monitoring.