Accuracy of an autonomous dental implant robotic system in partial edentulism: A pilot clinical study.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wang MiaoZhen, Liu Feng, Yu Tao, Zhan Yalin, Ma Feifei, Rausch-Fan Xiaohui
Аннотация
OBJECTIVES: Robots are increasingly being used for surgical procedures in various specialties. However, information about the accuracy of robot-assisted dental implant surgery is lacking. This pilot clinical study aimed to investigate the accuracy of an autonomous dental implant robotic (ADIR) system in partially edentulous cases.
MATERIAL AND METHODS: The ADIR system was used to place a total of 20 implants in 13 participants. Implant deviation from the planned positions was assessed to determine accuracy. The entry, apex, and angular deviations were described as means ± standard deviation. A two-sample t test was used to compare implant deviation between the flap and flapless groups and between maxillary and mandibular implants (α = .05).
RESULTS: The entry, apex, and angular deviations were 0.65 ± 0.32 mm, 0.66 ± 0.34 mm, and 1.52 ± 1.01°, respectively, with no statistically significant difference between the flap and flapless approaches (P > .05). No adverse events were encountered in any of the participants.
CONCLUSIONS: DIR accuracy in this clinical series was comparable to that reported for static and dynamic computer-assisted implant surgery. Robotic computer-assisted implant surgery may be useful for dental implant placement, potentially improving the quality and safety of the procedure.
CLINICAL RELEVANCE: The findings of this study showed that the ADIR system could be useful for dental implant surgery.
MATERIAL AND METHODS: The ADIR system was used to place a total of 20 implants in 13 participants. Implant deviation from the planned positions was assessed to determine accuracy. The entry, apex, and angular deviations were described as means ± standard deviation. A two-sample t test was used to compare implant deviation between the flap and flapless groups and between maxillary and mandibular implants (α = .05).
RESULTS: The entry, apex, and angular deviations were 0.65 ± 0.32 mm, 0.66 ± 0.34 mm, and 1.52 ± 1.01°, respectively, with no statistically significant difference between the flap and flapless approaches (P > .05). No adverse events were encountered in any of the participants.
CONCLUSIONS: DIR accuracy in this clinical series was comparable to that reported for static and dynamic computer-assisted implant surgery. Robotic computer-assisted implant surgery may be useful for dental implant placement, potentially improving the quality and safety of the procedure.
CLINICAL RELEVANCE: The findings of this study showed that the ADIR system could be useful for dental implant surgery.