European journal of heart failure · 2021
One-year results of the first-in-man study investigating the Atrial Flow Regulator for left atrial shunting in symptomatic heart failure patients: the PRELIEVE study.
СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Paitazoglou Christina, Bergmann Martin W, Özdemir Ramazan, Pfister Roman, Bartunek Jozef, Kilic Teoman, Lauten Alexander, Schmeisser Alexander, Zoghi Mehdi, Anker Stefan D, Sievert Horst, Mahfoud Felix, AFR-PRELIEVE Investigators
Аннотация
AIMS: Attenuating exercise-induced elevated left atrial pressure with an atrial shunt device is under clinical investigation for treatment of symptomatic heart failure (HF).
METHODS AND RESULTS: PRELIEVE was a prospective, non-randomised, multicentre, first-in-man study in symptomatic HF patients with reduced (HFrEF) or preserved (HFpEF) ejection fraction and pulmonary capillary wedge pressure (PCWP) ≥15 mmHg at rest or ≥25 mmHg during exercise. Here, we provide follow-up data up to 1 year after implantation of the Atrial Flow Regulator (AFR) device. The AFR was successfully implanted in 53 patients (HFrEF n = 24 and HFpEF n = 29). Two patients were not enrolled due to an unsuccessful transseptal puncture. There was one device embolisation into the left atrium, which required surgical removal. One patient experienced a serious procedure-related adverse event (post-procedural bleeding and syncope). All patients with sufficient echocardiography readout confirmed device patency with left-right shunt both at 3 (n = 47/51, 92%) and 12 (n = 45/49, 92%) months. At 3 months, rest PCWP decreased by 5 (-12, 0) mmHg (P = 0.0003, median Q1, Q3). No patient developed a stroke, worsening of right heart function or significant increase of pulmonary artery pressure. Six (6/53, 11%) patients were hospitalised for worsening of HF and three (3/53, 6%) patients died. We observed improvements in New York Heart Association functional class, 6-min walking distance and quality of life (Kansas City Cardiomyopathy Questionnaire) in certain patients.
CONCLUSIONS: Implantation of the AFR device in HF patients was feasible. No shunt occlusion, stroke or new right HF was observed during the 1-year follow-up, with clinical improvements in certain patients.
METHODS AND RESULTS: PRELIEVE was a prospective, non-randomised, multicentre, first-in-man study in symptomatic HF patients with reduced (HFrEF) or preserved (HFpEF) ejection fraction and pulmonary capillary wedge pressure (PCWP) ≥15 mmHg at rest or ≥25 mmHg during exercise. Here, we provide follow-up data up to 1 year after implantation of the Atrial Flow Regulator (AFR) device. The AFR was successfully implanted in 53 patients (HFrEF n = 24 and HFpEF n = 29). Two patients were not enrolled due to an unsuccessful transseptal puncture. There was one device embolisation into the left atrium, which required surgical removal. One patient experienced a serious procedure-related adverse event (post-procedural bleeding and syncope). All patients with sufficient echocardiography readout confirmed device patency with left-right shunt both at 3 (n = 47/51, 92%) and 12 (n = 45/49, 92%) months. At 3 months, rest PCWP decreased by 5 (-12, 0) mmHg (P = 0.0003, median Q1, Q3). No patient developed a stroke, worsening of right heart function or significant increase of pulmonary artery pressure. Six (6/53, 11%) patients were hospitalised for worsening of HF and three (3/53, 6%) patients died. We observed improvements in New York Heart Association functional class, 6-min walking distance and quality of life (Kansas City Cardiomyopathy Questionnaire) in certain patients.
CONCLUSIONS: Implantation of the AFR device in HF patients was feasible. No shunt occlusion, stroke or new right HF was observed during the 1-year follow-up, with clinical improvements in certain patients.