One-year results of the first-in-man study investigating the Atrial Flow Regulator for left atrial shunting in symptomatic heart failure patients: the PRELIEVE study.

Paitazoglou Christina, Bergmann Martin W, Özdemir Ramazan, Pfister Roman, Bartunek Jozef, Kilic Teoman, Lauten Alexander, Schmeisser Alexander, Zoghi Mehdi, Anker Stefan D, Sievert Horst, Mahfoud Felix, AFR-PRELIEVE Investigators

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: Attenuating exercise-induced elevated left atrial pressure with an atrial shunt device is under clinical investigation for treatment of symptomatic heart failure (HF).

METHODS AND

RESULTS

PRELIEVE was a prospective, non-randomised, multicentre, first-in-man study in symptomatic HF patients with reduced (HFrEF) or preserved (HFpEF) ejection fraction and pulmonary capillary wedge pressure (PCWP) ≥15 mmHg at rest or ≥25 mmHg during exercise. Here, we provide follow-up data up to 1 year after implantation of the Atrial Flow Regulator (AFR) device. The AFR was successfully implanted in 53 patients (HFrEF n = 24 and HFpEF n = 29). Two patients were not enrolled due to an unsuccessful transseptal puncture. There was one device embolisation into the left atrium, which required surgical removal. One patient experienced a serious procedure-related adverse event (post-procedural bleeding and syncope). All patients with sufficient echocardiography readout confirmed device patency with left-right shunt both at 3 (n = 47/51, 92%) and 12 (n = 45/49, 92%) months. At 3 months, rest PCWP decreased by 5 (-12, 0) mmHg (P = 0.0003, median Q1, Q3). No patient developed a stroke, worsening of right heart function or significant increase of pulmonary artery pressure. Six (6/53, 11%) patients were hospitalised for worsening of HF and three (3/53, 6%) patients died. We observed improvements in New York Heart Association functional class, 6-min walking distance and quality of life (Kansas City Cardiomyopathy Questionnaire) in certain patients.

CONCLUSIONS

Implantation of the AFR device in HF patients was feasible. No shunt occlusion, stroke or new right HF was observed during the 1-year follow-up, with clinical improvements in certain patients.

Для цитирования: Paitazoglou Christina, Bergmann Martin W, Özdemir Ramazan, Pfister Roman, Bartunek Jozef, Kilic Teoman, Lauten Alexander, Schmeisser Alexander, Zoghi Mehdi, Anker Stefan D, Sievert Horst, Mahfoud Felix, AFR-PRELIEVE Investigators. One-year results of the first-in-man study investigating the Atrial Flow Regulator for left atrial shunting in symptomatic heart failure patients: the PRELIEVE study.. European journal of heart failure. 2021. DOI: 10.1002/ejhf.2119.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.