DOI 10.1111/jce.14903

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2021

A new transseptal solution for enabling left atrial access of large delivery sheaths.

Inohara Taku, Gilhofer Thomas, Al-Dujaili Saja, Leung Linus, Yeung Darwin, Tsang Michael, Saw Jacqueline

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Transseptal access for large sheaths may be encumbered by tissue resistance against the sheath-dilator stepped interface. The ExpanSure Large Access Transseptal Dilator (Baylis Medical) is designed as a single introducer and dilation device with a smooth sheath-dilator transition to support transseptal puncture. It may facilitate ease and efficiency of interatrial crossing.

METHODS

This study experimentally evaluated the crossing force of ExpanSure relative to a conventional 8.5 F Swartz SL1 transseptal sheath and dilator in a benchtop septum model. Its ability to reduce the subsequent crossing force of a 14 F WATCHMAN delivery sheath was also tested. The clinical use of ExpanSure, including procedure time, was then validated in a series of left atrial appendage closure (LAAC) procedures.

RESULTS

In a benchtop septum model (N = 12), less peak force (1.90 ± 0.08 N vs. 2.36 ± 0.09 N; p < .001) and overall work (17.3 ± 1.2 mJ vs. 28.0 ± 1.9 mJ; p < .001) were required to advance ExpanSure relative to a conventional SL1 transseptal sheath and dilator system. Peak force (2.34 ± 0.24 N vs. 2.65 ± 0.21 N; p < .003) and overall work (28.5 ± 3.9 mJ vs. 35.4 ± 2.1 mJ; p < .001) to advance a WATCHMAN sheath were also significantly lower after using ExpanSure than after using a conventional transseptal system. In 19 LAAC procedures, ExpanSure crossed the septum smoothly and integrated readily, which enabled efficient procedure completion (mean total procedure time 37.6 ± 13.5 min), with 100% success and no procedure-related complications.

CONCLUSION

Experimental force measurements, combined with early clinical experience using ExpanSure, suggest that the tapered design with smooth transition without dilator-sheath step-up and the larger diameter, both facilitated ease and efficiency of interatrial crossing.

Для цитирования: Inohara Taku, Gilhofer Thomas, Al-Dujaili Saja, Leung Linus, Yeung Darwin, Tsang Michael, Saw Jacqueline. A new transseptal solution for enabling left atrial access of large delivery sheaths.. Journal of cardiovascular electrophysiology. 2021. DOI: 10.1111/jce.14903.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.