MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jce.14824

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2020

Incidence, risk factors, and clinical impact of peridevice leak following left atrial appendage closure with the LAmbre device-Data from a prospective multicenter clinical study.

Wang Guangji, Kong Bin, Qin Tianyou, Liu Yu, Huang Congxin, Huang He

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

In the present study, we sought to explore the incidence, risk factors, and clinical impact of peridevice leaks (PDLs), following LAmbre-assisted left atrial appendage closure (LAAC).

METHODS

We performed transesophageal echocardiography (TEE) on patients participating in the LAmbre multicenter study, at Day 1 postimplantation, then at 3 and 12 months to assess PDL, device-related thrombus, left atrial appendage (LAA) thrombus, and left atrial thrombus. Clinical events were recorded during follow-up.

RESULT: A total of 152 patients with atrial fibrillation successfully completed LAAC. At 3 months follow-up, 123 patients underwent TEE, with 21 (17%) of them presenting PDL. Among the 121 patients who underwent TEE at 12 months follow-up, 19 (15.7%) presented PDL. Patients with PDL exhibited larger LAA orifice diameters and larger device sizes compared to those in the no leak group. In addition, we found no significant differences in thromboembolic events between patients in the PDL and no leak groups.

CONCLUSION

LAmbre-assisted LAA closure resulted in a relatively low PDL occurrence, and its rate decreased over time. In addition, PDL was more prominent in patients with larger LAA orifice diameter and larger device size. However, the condition was not associated with an increased risk for thromboembolic events.

Для цитирования: Wang Guangji, Kong Bin, Qin Tianyou, Liu Yu, Huang Congxin, Huang He. Incidence, risk factors, and clinical impact of peridevice leak following left atrial appendage closure with the LAmbre device-Data from a prospective multicenter clinical study.. Journal of cardiovascular electrophysiology. 2020. DOI: 10.1111/jce.14824.