Hickey Samantha, Reichert Andreas, Ptacek Wolfgang, Bielak Lars, Reiss Simon, Fischer Johannes, Gunashekar Deepa Darshini, Bortfeld Thomas, Bock Michael
Аннотация
PURPOSE: MR guidance is used during therapy to detect and compensate for lesion motion. T2 -weighted MRI often has a superior lesion contrast in comparison to T1 -weighted real-time imaging. The purpose of this work was to design a fast T2 -weighted sequence capable of simultaneously acquiring two orthogonal slices, enabling real-time tracking of lesions.
METHODS: To generate a T2 contrast in two orthogonal slices simultaneously, a sequence (Ortho-SFFP-Echo) was designed that samples the T2 -weighted spin echo (S- ) signal in a TR-interleaved acquisition of two slices. Slice selection and phase-encoding directions are swapped between the slices, leading to a unique set of spin-echo signal conditions. To minimize motion-related signal dephasing, additional flow-compensation strategies are implemented. In both the abdominal breathing phantom and in vivo experiments, a time series was acquired using Ortho-SSFP-Echo. The centroid of the target was tracked in postprocessing steps.
RESULTS: In the phantom, the lesion could be identified and delineated in the dynamic images. In the volunteer experiments, the kidney was visualized with a T2 contrast at a temporal resolution of 0.45 s under free-breathing conditions. A respiratory belt demonstrated a strong correlation with the time course of the kidney centroid in the head-foot direction. A hypointense saturation band at the slice overlap did not inhibit lesion tracking in the semi-automatic postprocessing steps.
CONCLUSION: The Ortho-SFFP-Echo sequence delivers real-time images with a T2 -weighted contrast in two orthogonal slices. The sequence allows for simultaneous acquisition, which could be beneficial for real-time motion tracking in radiotherapy or interventional MRI.
METHODS: To generate a T2 contrast in two orthogonal slices simultaneously, a sequence (Ortho-SFFP-Echo) was designed that samples the T2 -weighted spin echo (S- ) signal in a TR-interleaved acquisition of two slices. Slice selection and phase-encoding directions are swapped between the slices, leading to a unique set of spin-echo signal conditions. To minimize motion-related signal dephasing, additional flow-compensation strategies are implemented. In both the abdominal breathing phantom and in vivo experiments, a time series was acquired using Ortho-SSFP-Echo. The centroid of the target was tracked in postprocessing steps.
RESULTS: In the phantom, the lesion could be identified and delineated in the dynamic images. In the volunteer experiments, the kidney was visualized with a T2 contrast at a temporal resolution of 0.45 s under free-breathing conditions. A respiratory belt demonstrated a strong correlation with the time course of the kidney centroid in the head-foot direction. A hypointense saturation band at the slice overlap did not inhibit lesion tracking in the semi-automatic postprocessing steps.
CONCLUSION: The Ortho-SFFP-Echo sequence delivers real-time images with a T2 -weighted contrast in two orthogonal slices. The sequence allows for simultaneous acquisition, which could be beneficial for real-time motion tracking in radiotherapy or interventional MRI.