Медицинская библиотека · Diabetes, obesity & metabolism · 2025

Accuracy and reliability of a continuous glucose monitoring system with a focus on hypoglycaemia.

van Baal Lukas, Raess Lorenz, Mathew Annie, Lahner Harald, Unger Nicole, Fuhrer Dagmar

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Glucose monitoring is increasingly based on continuous glucose monitoring (CGM) systems. However, data on the accuracy of CGM within the hypoglycaemic range is sparse. This study investigated CGM accuracy within the hypoglycaemic range in three different clinical settings of hypoglycaemia.

MATERIALS AND

METHODS

Ninety-two people with various causes of hypoglycaemia were analysed: (i) people during insulin tolerance testing (ITT) (n = 63); (ii) people with insulinoma (n = 16); and (iii) people with diabetes receiving subcutaneous insulin therapy (n = 13). CGM accuracy was evaluated for subgroups and different glucose rates of change (RoC) using mean absolute relative difference (MARD), percentage of glucose values within ±20 mg/dL of point-of-care glucose (%20/20), diabetes technology society error grid and Bland-Altman analysis (BAA).

RESULTS

Four hundred sixty-four CGM/POC glucose pairs were obtained (39.7% level 1, 60.3% level 2 hypoglycaemia). CGM accuracy decreased from people with diabetes receiving subcutaneous insulin therapy (MARD: 13.9%; %20/20: 81.5%) to those with ITT (MARD: 50.8%; p < 0.01, %20/20: 67.5%; p .02). CGM accuracy significantly decreased with higher RoC. Proportion of CGM values with moderate risk in failing to detect potentially dangerous hypoglycaemia increased from people with diabetes (1.5%) and insulinoma (1.2%) to people with ITT (14.2%, p < 0.01). BAA revealed a significantly increasing bias of -3.3 + 10.7 mg/dL in people with diabetes receiving insulin therapy to -15.2 ± 13.6 mg/dL in people with ITT (p < 0.01).

CONCLUSION

CGM accuracy can vary in different clinical hypoglycaemia scenarios. While it remains acceptable in people with diabetes receiving subcutaneous insulin therapy and people with insulinoma, it appears to be inaccurate for glucose monitoring during ITT.

Для цитирования: van Baal Lukas, Raess Lorenz, Mathew Annie, Lahner Harald, Unger Nicole, Fuhrer Dagmar. Accuracy and reliability of a continuous glucose monitoring system with a focus on hypoglycaemia.. Diabetes, obesity & metabolism. 2025. DOI: 10.1111/dom.70408.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.