Orthodontics & craniofacial research · 2013

Accuracy of volumetric measurement of simulated root resorption lacunas based on cone beam computed tomography.

Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Wang Y, He S, Guo Y, Wang S, Chen S

Аннотация

OBJECTIVES: To evaluate the accuracy of volumetric measurement of simulated root resorption cavities based on cone beam computed tomography (CBCT), in comparison with that of Micro-computed tomography (Micro-CT) which served as the reference.

SETTING AND SAMPLE POPULATION: The State Key Laboratory of Oral Diseases at Sichuan University.

MATERIAL AND METHODS: Thirty-two bovine teeth were included for standardized CBCT scanning and Micro-CT scanning before and after the simulation of different degrees of root resorption. The teeth were divided into three groups according to the depths of the root resorption cavity (group 1: 0.15, 0.2, 0.3 mm; group 2: 0.6, 1.0 mm; group 3: 1.5, 2.0, 3.0 mm). Each depth included four specimens. Differences in tooth volume before and after simulated root resorption were then calculated from CBCT and Micro-CT scans, respectively. The overall between-method agreement of the measurements was evaluated using the concordance correlation coefficient (CCC).

RESULTS: For the first group, the average volume of resorption cavity was 1.07 mm(3) , and the between-method agreement of measurement for the volume changes was low (CCC = 0.098). For the second and third groups, the average volumes of resorption cavities were 3.47 and 6.73 mm(3) respectively, and the between-method agreements were good (CCC = 0.828 and 0.895, respectively).

CONCLUSIONS: The accuracy of 3-D quantitative volumetric measurement of simulated root resorption based on CBCT was fairly good in detecting simulated resorption cavities larger than 3.47 mm(3), while it was not sufficient for measuring resorption cavities smaller than 1.07 mm(3) . This method could be applied in future studies of root resorption although further studies are required to improve its accuracy.