Quantification of external root resorption by low- vs high-resolution cone-beam computed tomography and periapical radiography: A volumetric and linear analysis.

Ponder Stacy N, Benavides Erika, Kapila Sunil, Hatch Nan E

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The goal of this study was to determine whether cone beam-computed tomography (CBCT) images with resolutions similar to those produced in orthodontic offices have sufficient resolution to accurately quantify root resorption defects.

METHODS

Teeth containing simulated root defects were scanned by microcomputed tomography (microCT) and CBCT at 0.2- and 0.4-mm resolutions and were radiographed by the periapical technique. Root length was measured with digital calipers. Comparisons were made to establish significance between imagining modalities and to compare defects of differing severity and position.

RESULTS

The mean absolute difference in volumetric measurements of lateral defects from 0.2-mm-resolution CBCT images compared with those from microCT images was significantly smaller than those from 0.4-mm-resolution CBCT images (0.20 ± 0.2 mm(3) vs 0.30 ± 0.3 mm(3); P = 0.002). A Bland-Altman analysis showed that the 95% limits of agreement range between low-resolution CBCT and microCT volumetric measurements was1.44-fold greater than that between high-resolution CBCT and microCT measurements (-0.87-0.68 vs -0.49-0.59 mm(3)). The accuracy of the volumetric measurements was also significantly influenced by defect size (P = 0.004 for high-resolution CBCT, P = 0.005 for low-resolution CBCT) and, on low-resolution scans, by the defect's vertical position (P = 0.012). Linear measurements of apical defects from both the 0.2- and 0.4-mm-resolution CBCT images were significantly more similar to the measurements made with digital calipers than the measurements from the periapical radiographs (P <0.0001 and P <0.0001, respectively).

CONCLUSIONS

Our results demonstrated that, compared with measurements from microCT images, high-resolution CBCT scans lead to more accurate volumetric quantifications of lateral resorption defects than do low-resolution scans. Both high- and low-resolution CBCT scans can also be used to more accurately measure external apical root resorption defects than periapical radiographs. Because these results are from CBCT scans of static images, measurements of root resorption defects from scans of patients will most likely be less accurate because of patient movement.

Для цитирования: Ponder Stacy N, Benavides Erika, Kapila Sunil, Hatch Nan E. Quantification of external root resorption by low- vs high-resolution cone-beam computed tomography and periapical radiography: A volumetric and linear analysis.. American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics. 2013. DOI: 10.1016/j.ajodo.2012.08.023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.