Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years.

Kleemann Thomas, Becker Torsten, Doenges Klaus, Vater Margit, Senges Jochen, Schneider Steffen, Saggau Werner, Weisse Udo, Seidl Karlheinz

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The number of patients with longer follow-up after implantation of an implantable cardioverter-defibrillator is increasing continuously. Defibrillation lead failure is a typical long-term complication. Therefore, the long-term reliability of implantable cardioverter-defibrillator leads has become an increasing concern. The aim of the present study was to assess the annual rate of transvenous defibrillation lead defects related to follow-up time after lead implantation.

METHODS AND

RESULTS

A total of 990 consecutive patients who underwent first implantation of an implantable cardioverter-defibrillator between 1992 and May 2005 were analyzed. Median follow-up time was 934 days (interquartile range, 368 to 1870). Overall, 148 defibrillation leads (15%) failed during the follow-up. The estimated lead survival rates at 5 and 8 years after implantation were 85% and 60%, respectively. The annual failure rate increased progressively with time after implantation and reached 20% in 10-year-old leads (P<0.001). Lead defects affected newer as well as older models. Patients with lead defects were 3 years younger at implantation and more often female. Multiple lead implantation was associated with a trend to a higher rate of defibrillation lead defects (P=0.06). The major lead complications were insulation defects (56%), lead fractures (12%), loss of ventricular capture (11%), abnormal lead impedance (10%), and sensing failure (10%).

CONCLUSIONS

An increasing annual lead failure rate is noted primarily during long-term follow-up and reached 20% in 10-year-old leads. Patients with lead defects are younger and more often female.

Для цитирования: Kleemann Thomas, Becker Torsten, Doenges Klaus, Vater Margit, Senges Jochen, Schneider Steffen, Saggau Werner, Weisse Udo, Seidl Karlheinz. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years.. Circulation. 2007. DOI: 10.1161/circulationaha.106.663807.