Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator.

Danik Stephan B, Mansour Moussa, Singh Jagmeet, Reddy Vivek Y, Ellinor Patrick T, Milan David, Heist E Kevin, d'Avila Andre, Ruskin Jeremy N, Mela Theofanie

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The rapid evolution of implantable cardioverter-defibrillator (ICD) leads has resulted in thinner active fixation leads. While these advances have made the leads more versatile, new configurations may be associated with unforeseen complications.

OBJECTIVE

The purpose of this study was to determine the incidence of perforation and dislodgement of defibrillator leads in a single center in the year 2005.

METHODS

All patients who underwent percutaneous ICD implantation at the Massachusetts General Hospital using an endocardial right ventricular lead were included in this study. The specific leads analyzed were the Riata (1580/1581 and 1590/1591, St. Jude Medical, St Paul, Minnesota, USA;) and Sprint Fidelis (6949-65, Medtronic, Minneapolis, Minnesota, USA.). Information was collected retrospectively.

RESULTS

A total of 130 Riata leads and 111 Sprint Fidelis leads were implanted at the Massachusetts General Hospital during this time period. A total of five lead perforations occurred in patients implanted with the Riata lead as compared with none with the Sprint Fidelis lead (3.8% vs. 0%, respectively; P <.05). Two of the five patients with perforation required pericardiocentesis for tamponade. Clinical symptoms of perforation developed 1-10 days after implant. Moreover, there were five additional lead revisions in the Riata group, which were likely due to dislodgement and/or microperforation, as compared with none in the Sprint Fidelis group (7.7% vs. 0%, respectively; P <.005).

CONCLUSIONS

In 2005, at one institution, there were significantly more cardiac perforations and lead revisions with the Riata lead as compared with the Sprint Fidelis right ventricular defibrillator lead. Further data are required to determine whether certain lead characteristics are responsible for this observation.

Для цитирования: Danik Stephan B, Mansour Moussa, Singh Jagmeet, Reddy Vivek Y, Ellinor Patrick T, Milan David, Heist E Kevin, d'Avila Andre, Ruskin Jeremy N, Mela Theofanie. Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator.. Heart rhythm. 2007. DOI: 10.1016/j.hrthm.2006.12.044.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.