MedElevator Войти в исследования
← Поиск публикаций

DOI 10.4103/0974-2700.115340

Медицинская библиотека · Journal of emergencies, trauma, and shock · 2013

Bedside ultrasonography by emergency physicians for anterior talofibular ligament injury.

Gün Cem, Unlüer Erden Erol, Vandenberk Nergiz, Karagöz Arif, Sentürk Güldehen Ozmen, Oyar Orhan

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Our objective was to study the accuracy of emergency physician (EP) performed bedside ultrasonography (BUS) in patients with suspected anterior talofibular ligament (ATFL) injury.

MATERIALS AND

METHODS

After a 6-h training program, from January to December 2011, an EP used BUS to prospectively evaluate patients presenting to the emergency department (ED) with suspected ATFL injury. Then, patients underwent ankle X-ray and Magnetic Resonance (MR) imaging. Outcome was determined by official radiology reports of the MR imaging. BUS and MR imaging results were compared using Chi-square testing.

RESULTS

Of the 65 enrolled patients, 30 patients were BUS positive. Of these, MR imaging results agreed with the BUS findings in 30 patients. In 35 cases, BUS was negative, and 33 of these were corroborated by MR imaging. The sensitivity, specificity, positive predictive value, negative predictive value, and negative likelihood ratio for BUS were 93.8%, 100%, 100%, 94.3%, and 0.06%, respectively. The diagnostic accuracy of BUS was not statistically different from MR imaging (K = 0.938, P = 0.001).

CONCLUSION

BUS for the diagnosis of ATFL injury is another application of BUS in the ED. EPs can diagnose ATFL injury using BUS with a high degree of accuracy.

Для цитирования: Gün Cem, Unlüer Erden Erol, Vandenberk Nergiz, Karagöz Arif, Sentürk Güldehen Ozmen, Oyar Orhan. Bedside ultrasonography by emergency physicians for anterior talofibular ligament injury.. Journal of emergencies, trauma, and shock. 2013. DOI: 10.4103/0974-2700.115340.