Reliability of the Watch-PAT 200 in detecting sleep apnea in highway bus drivers.

Yuceege Melike, Firat Hikmet, Demir Ahmet, Ardic Sadik

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To predict the validity of Watch-PAT (WP) device for sleep disordered breathing (SDB) among highway bus drivers.

METHOD: A total number of 90 highway bus drivers have undergone polysomnography (PSG) and Watch-PAT test simultaneously. Routine blood tests and the routine ear-nose-throat (ENT) exams have been done as well.

RESULTS

The sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were 89.1%, 76.9%, 82% and 85.7% for RDI > 15, respectively. WRDI, WODI, W < 90% duration and Wmean SaO2 results were well correlated with the PSG results. In the sensitivity and specificity analysis, when diagnosis of sleep apnea was defined for different cut-off values of RDI of 5, 10 and 15, AUC (95%CI) were found as 0.84 (0.74-0.93), 0.87 (95%CI: 0.79-0.94) and 0.91 (95%CI: 0.85-0.97), respectively. There were no statistically significant differences between Stage1+2/Wlight and Stage REM/WREM. The percentage of Stage 3 sleep had difference significant statistically from the percentage of Wdeep. Total sleep times in PSG and WP showed no statistically important difference. Total NREM duration and total WNREM duration had no difference either.

CONCLUSION

Watch-PAT device is helpful in detecting SDB with RDI > 15 in highway bus drivers, especially in drivers older than 45 years, but has limited value in drivers younger than 45 years old who have less risk for OSA. Therefore, WP can be used in the former group when PSG is not easily available.

Для цитирования: Yuceege Melike, Firat Hikmet, Demir Ahmet, Ardic Sadik. Reliability of the Watch-PAT 200 in detecting sleep apnea in highway bus drivers.. Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine. 2013. DOI: 10.5664/jcsm.2584.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.