Radiation Exposure of Interventional Radiologists during Computed Tomography Fluoroscopy-Guided Percutaneous Cryoablation.

Seki Yuko, Miyazaki Masaya, Fukushima Yasuhiro, Ando Masashi, Tsushima Yoshito

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The principal aim of this study was to evaluate radiation exposure of interventionalists during computed tomography (CT) fluoroscopy-guided percutaneous cryoablation (PCA) using radiophotoluminescent glass dosimeters (RPLDs). The radioprotective effects of safety glasses and lead apron were also evaluated.

MATERIALS AND

METHODS

Radiation exposure of interventionalists during 46 CT fluoroscopy-guided PCA procedures was evaluated. Entrance surface dose (ESD) was measured using RPLDs on multiple sites: five sites, representing eye lens exposure; five sites, representing body exposure; and four sites, representing skin exposure. The ESD values on multiple sites were compared between different PCA procedures (renal, liver, and bone).

RESULTS

The mean ESD on the X-ray-side hand exhibited the highest value (358.8 μGy). Regarding evaluation sites representing exposure to the eye lens, the highest ESD inside the radiation protective glasses was detected on the X-ray-side cheek (167.1 μGy). Most ESD values among multiple sites (10/14) were linearly correlated with CT fluoroscopy time. Among them, the ESD values measured during renal and liver PCA were relatively higher than those measured during bone PCA, especially on the chest area outside the lead apron, and on the X-ray tube-side elbow and hand during renal and bone PCA. Radioprotective effects of safety glasses and lead apron ranged from 44.6 to 50.6% and from 30.2 to 79.6%, respectively, on each evaluation site.

CONCLUSION

The site with the highest radiation exposure on interventionalists during CT fluoroscopy-guided PCA was the X-ray tube-side hand. Radiation exposure of interventionalists was at acceptable levels and consistent with the recommended dose limits.

Для цитирования: Seki Yuko, Miyazaki Masaya, Fukushima Yasuhiro, Ando Masashi, Tsushima Yoshito. Radiation Exposure of Interventional Radiologists during Computed Tomography Fluoroscopy-Guided Percutaneous Cryoablation.. Interventional radiology (Higashimatsuyama-shi (Japan). 2020. DOI: 10.22575/interventionalradiology.2019-0009.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.