Journal of clinical laboratory analysis · 2021
Diagnostic performance of the Elecsys SARS-CoV-2 antigen assay in the clinical routine of a tertiary care hospital: Preliminary results from a single-center evaluation.
Оценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mueller Thomas, Kompatscher Julia, La Guardia Mario
Аннотация
BACKGROUND: This report describes a manufacturer-independent evaluation of the diagnostic accuracy of the Elecsys SARS-CoV-2 antigen assay from Roche Diagnostics in a tertiary care setting.
METHODS: In this single-center study, we used nasopharyngeal swabs from 403 cases from the emergency department and intensive care unit of our hospital. The reference standard for detecting SARS-CoV-2 was the reverse-transcription polymerase chain reaction (RT-PCR) assay. Cycle threshold (Ct) values were recorded for positive RT-PCR assays. The index test was the Elecsys SARS-CoV-2 antigen assay. This electrochemiluminescence immunoassay produces results as cutoff index (COI) values, with values ≥1.00 being reported as positive.
RESULTS: Of the 403 cases, 47 showed positive results in RT-PCR assays. Of the 47 RT-PCR-positive cases, 12 showed positive results in the antigen assay. Of the 356 RT-PCR-negative cases, all showed negative results in the antigen assay. Thus, the antigen assay showed a sensitivity of 26% (95% CI, 14%-40%) and specificity of 100% (95% CI, 99%-100%). Analysis of the relationship between Ct values and COI values in the 47 RT-PCR-positive cases showed a correlation coefficient of -0.704 (95% CI, -0.824 to -0.522). The true-positive rate of the antigen assay for Ct values of 15-24.9, 25-29.9, 30-34.9, and 35-39.9 was 100%, 44%, 8%, and 6%, respectively.
CONCLUSIONS: The Elecsys SARS-CoV-2 antigen assay has a low sensitivity for detecting SARS-CoV-2 from nasopharyngeal swabs. Hence, we decided to not use this assay in the clinical routine of our hospital.
METHODS: In this single-center study, we used nasopharyngeal swabs from 403 cases from the emergency department and intensive care unit of our hospital. The reference standard for detecting SARS-CoV-2 was the reverse-transcription polymerase chain reaction (RT-PCR) assay. Cycle threshold (Ct) values were recorded for positive RT-PCR assays. The index test was the Elecsys SARS-CoV-2 antigen assay. This electrochemiluminescence immunoassay produces results as cutoff index (COI) values, with values ≥1.00 being reported as positive.
RESULTS: Of the 403 cases, 47 showed positive results in RT-PCR assays. Of the 47 RT-PCR-positive cases, 12 showed positive results in the antigen assay. Of the 356 RT-PCR-negative cases, all showed negative results in the antigen assay. Thus, the antigen assay showed a sensitivity of 26% (95% CI, 14%-40%) and specificity of 100% (95% CI, 99%-100%). Analysis of the relationship between Ct values and COI values in the 47 RT-PCR-positive cases showed a correlation coefficient of -0.704 (95% CI, -0.824 to -0.522). The true-positive rate of the antigen assay for Ct values of 15-24.9, 25-29.9, 30-34.9, and 35-39.9 was 100%, 44%, 8%, and 6%, respectively.
CONCLUSIONS: The Elecsys SARS-CoV-2 antigen assay has a low sensitivity for detecting SARS-CoV-2 from nasopharyngeal swabs. Hence, we decided to not use this assay in the clinical routine of our hospital.