DOI 10.1016/j.jiac.2021.06.019

Медицинская библиотека · Journal of infection and chemotherapy : official journal of the Japan Society of Chemotherapy · 2021

Evaluation of false positives in the SARS-CoV-2 quantitative antigen test.

Kobayashi Ryo, Murai Ryosei, Moriai Mikako, Nirasawa Shinya, Yonezawa Hitoshi, Kondoh Takashi, Saeki Masachika, Yakuwa Yuki, Sato Yuki, Katayama Yuki, Nakafuri Hirotaka, Kitayama Ikumi, Asanuma Koichi, Fujiya Yoshihiro, Takahashi Satoshi

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Highly sensitive reagents for detecting SARS-CoV-2 antigens have been developed for accurate and rapid diagnosis till date. In this study, we aim to clarify the frequency of false-positive reactions and reveal their details in SARS-CoV-2 quantitative antigen test using an automated laboratory device.

METHODS

Nasopharyngeal swab samples (n = 4992) and saliva samples (n = 5430) were collected. We measured their SARS-CoV-2 antigen using Lumipulse® Presto SARS-CoV-2 Ag and performed a nucleic acid amplification test (NAAT) using the Ampdirect™ 2019 Novel Coronavirus Detection Kit as needed. The results obtained from each detection test were compared accordingly.

RESULTS

There were 304 nasopharyngeal samples and 114 saliva samples were positive in the Lumipulse® Presto SARS-CoV-2 Ag test. All positive nasopharyngeal samples in the antigen test were also positive for NAAT. In contrast, only three (2.6%) of all the positive saliva samples in the antigen test were negative for NAAT. One showed no linearity with a dilute solution in the dilution test. Additionally, the quantitative antigen levels of all the three samples did not decrease after reaction with the anti-SARS-CoV-2 antibody.

CONCLUSIONS

The judgment difference between the quantitative antigen test and NAAT seemed to be caused by non-specific reactions in the antigen test. Although the high positive and negative predictive value of this quantitative antigen test could be confirmed, we should consider the possibility of false-positives caused by non-specific reactions and understand the characteristics of antigen testing. We recommend that repeating centrifugation before measurement, especially in saliva samples, should be performed appropriately.

Для цитирования: Kobayashi Ryo, Murai Ryosei, Moriai Mikako, Nirasawa Shinya, Yonezawa Hitoshi, Kondoh Takashi, Saeki Masachika, Yakuwa Yuki, Sato Yuki, Katayama Yuki, Nakafuri Hirotaka, Kitayama Ikumi, Asanuma Koichi, Fujiya Yoshihiro, Takahashi Satoshi. Evaluation of false positives in the SARS-CoV-2 quantitative antigen test.. Journal of infection and chemotherapy : official journal of the Japan Society of Chemotherapy. 2021. DOI: 10.1016/j.jiac.2021.06.019.