DOI 10.1016/j.ejrad.2025.112540
Optimizing photon-counting CT pulmonary angiography for robustness, combining mixed bolus, fixed scan delay and low keV.
Аннотация
PURPOSE: This study's purpose was to create a robust, optimized CTPA protocol by exploring different combinations of keV reconstructions and contrast infusion parameters.
MATERIALS &
METHODS
PCCT CTPAs between July 2023-January 2024 were retrospectively included, spanning three contrast injection protocols during a period of protocol optimization: cohort 1: 50/0mL contrast/saline mix using bolus tracking; cohort 2: 50/50 mL mix and cohort 3: 35/35 mL mix, both with fixed scan delay. Patients were matched by water-equivalent diameter between cohorts. Images were reconstructed at 40, 55 and 67 keV. Hounsfield unit (HU) values were measured in the main pulmonary artery and the segmental arteries. Three blinded radiologists rated subjective image quality.
RESULTS
150 patients (49 men, 101 women) were evaluated (mean age: 60.6 ± 16.6 years), 50 per cohort. Mean HU values were significantly higher in cohort 1 > 2 > 3 (p < 0.001) and were significantly higher at 40 > 55 > 67 keV (p < 0.001). At 40 keV, three (1.5 %) cohort 1 contrast injections were < 350 HU, with none < 350 HU in cohorts 2 or 3. Readers rated contrast best at 40 keV for vascular assessment (cohort 2: 4.7 ± 0.5, vs 4.5 ± 0.7 at 55 keV, vs 3.6 ± 1.0 at 67 keV, p < 0.001).
CONCLUSION
Our study suggests that the most robust PCCT contrast injection protocol for CTPA is a mixed injection of 50 or 35 mL of contrast with the same volume of saline, using a fixed scan delay. 40 keV vascular reconstructions are favored by radiologists and minimize non-diagnostic contrast injections.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.