Neonatal oesophageal and axilla temperatures in the neonatal intensive care unit care.
Аннотация
AIM: To compare oesophageal and axilla temperatures in routine neonatal intensive care unit (NICU) care.
METHODS
Prospective observational study of NICU infants with indwelling oral/nasogastric tubes. Three simultaneous temperature measurements were taken: Physitemp axilla, digital Premie Faichney axilla in predictive mode and Physitemp oesophageal. Temperatures were measured on two separate days.
RESULTS
Fifty infants were studied over a range of gestational ages (median 28.9 weeks, IQR 25.8-30.6 weeks) and birthweights (median 1140 g, IQR 742-1498 g). Mean Physitemp oesophageal temperatures were 0.3 ± 0.1°C (range 0.1-0.6°C) higher than Physitemp axilla temperatures. After adjustment for weight, the Physitemp oesophageal-Physitemp axilla difference remained constant over the average oesophageal-axilla temperature range studied (p = 0.07). Physitemp oesophageal temperatures were not affected by weight (p = 0.2) or postmenstrual age (p = 0.51). Physitemp axilla temperatures decreased with increasing weight (p = 0.03) and postmenstrual age (p = 0.048). The Physitemp oesophageal-Physitemp axilla difference was greater for infants in cribs (mean ± SD = 0.4 ± 0.1°C) than in incubators (mean ± SD = 0.3 ± 0.1°C, p < 0.001). The relationship between oesophageal and digital Premie Faichney axilla temperature was similar to oesophageal and axilla Physitemp temperatures.
CONCLUSION
Over a range of infant temperatures encountered in routine NICU care, oesophageal temperature is higher than axilla temperature, although the difference is small. Axillary temperatures measured by digital thermometers in predictive mode are comparable to core oesophageal temperatures.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.