Hasler William L, Abell Thomas, Attaluri Ashok, Chey William D, Cline Michael, Gearry Richard B, Gibson Peter R, Ho Vincent, Lee Allen A, Lembo Anthony J, Masoud Amir, McCallum Richard, Moshiree Baharak, Quigley Eamonn M M, Rao Satish S C, Sanchez Mayra, Sarosiek Irene, Stocker Abigail, Surjanhata Brian, Thwaites Phoebe A, Zhou Jerry, Kuo Braden
Аннотация
BACKGROUND: Patients with gastric dysmotility exhibit abnormal transit and reduced contractility, for which quantification is limited with current methodologies. The Atmo Capsule is a wireless motility capsule that measures regional gastrointestinal transit. We tested its technical capabilities to evaluate gastric contractility by detecting changes in capsule acceleration during transit.
METHODS: Data from 91 healthy controls and 182 patients with suspected dysmotility who ingested the Atmo Capsule following a standardized protocol were included. Gastric emptying times (GET) ≥ 5 h were considered delayed. Contraction frequencies (cycles per minute [cpm]) and amplitudes (as arbitrary units [au]) were identified following Fast Fourier Transformation of accelerometer data using 5-min windows and were assessed in 30-min windows post-ingestion (early gastric phase) and before gastric emptying (late gastric phase). Data were compared via ANOVA, using Bonferroni corrections, between healthy controls, patients with normal gastric emptying, and patients with delayed gastric emptying.
KEY RESULTS: Gastric emptying was delayed in 53/182 (29%) patients. Three cpm contraction frequencies were identified across both gastric phases in controls and patients. Contraction amplitudes did not differ between groups in the early gastric phase. In the late phase, mean amplitudes were lower in patients with delayed gastric emptying (6.4 au) compared with subjects with normal gastric emptying (controls, 9.7 au, p < 0.0001; patients, 8.6 au, p < 0.0001).
CONCLUSIONS & INFERENCE: The Atmo Capsule enables precise quantification of gastric contraction profiles, permitting differentiation of normal from pathological motor patterns. Findings of this technical assessment support its potential for evaluating gastric physiology and detecting pathophysiological patterns of gastric dysmotility.
METHODS: Data from 91 healthy controls and 182 patients with suspected dysmotility who ingested the Atmo Capsule following a standardized protocol were included. Gastric emptying times (GET) ≥ 5 h were considered delayed. Contraction frequencies (cycles per minute [cpm]) and amplitudes (as arbitrary units [au]) were identified following Fast Fourier Transformation of accelerometer data using 5-min windows and were assessed in 30-min windows post-ingestion (early gastric phase) and before gastric emptying (late gastric phase). Data were compared via ANOVA, using Bonferroni corrections, between healthy controls, patients with normal gastric emptying, and patients with delayed gastric emptying.
KEY RESULTS: Gastric emptying was delayed in 53/182 (29%) patients. Three cpm contraction frequencies were identified across both gastric phases in controls and patients. Contraction amplitudes did not differ between groups in the early gastric phase. In the late phase, mean amplitudes were lower in patients with delayed gastric emptying (6.4 au) compared with subjects with normal gastric emptying (controls, 9.7 au, p < 0.0001; patients, 8.6 au, p < 0.0001).
CONCLUSIONS & INFERENCE: The Atmo Capsule enables precise quantification of gastric contraction profiles, permitting differentiation of normal from pathological motor patterns. Findings of this technical assessment support its potential for evaluating gastric physiology and detecting pathophysiological patterns of gastric dysmotility.