m+Medelevator.RU
← Поиск публикаций
Medical physics · 2026

Dual-view scout scans with deep learning for ultra-low dose attenuation correction in PET.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Muller Florence M, Daube-Witherspoon Margaret E, Parma Michael J, Perkins Amy E, Noël Peter B, Vanhove Christian, Vandenberghe Stefaan, Karp Joel S

Аннотация

BACKGROUND: Accurate attenuation correction (AC) is essential for quantitative positron emission tomography (PET). Conventional CT-based AC provides reliable attenuation (μ-) maps but adds radiation, introduces PET/CT misalignment artifacts, and is unavailable on stand-alone PET systems. Existing deep learning (DL) methods use non-attenuation corrected (NAC) PET data to predict CT-like or directly AC-PET images, but their dependence on emission characteristics limits generalizability across tracers and anatomical coverage, especially for long axial field-of-view PET.

PURPOSE: We propose incorporating tissue-density information from dual-view scout radiographs as supplementary input to NAC PET, acquired with a clinically significant radiation reduction relative to standard low-dose CT.

METHODS: Two DL methods were evaluated for generating transmission (Tr) images as surrogate μ-maps for PET AC: (1) (NAC)-to-(Tr), using coronal and sagittal NAC PET slices, and (2) (NAC + Scout)-to-(Tr), combining NAC PET with anterior-posterior and lateral scout views. 53 research scans across four tracers from the PennPET Explorer were used for training and testing, with three additional tracers included for testing.

RESULTS: (NAC + Scout)-to-(Tr) improved quantitative accuracy, reducing NRMSE% to < 10% (versus up to 18% for (NAC)-to-(Tr)) and SUV biases to within -10%, with significant reductions in brain, liver, and muscle compared to (NAC)-to-(Tr). Out-of-distribution evaluation confirmed generalizability, maintaining SUV biases below 5%-8% versus 10%-13% for (NAC)-to-(Tr). In a longitudinal biodistribution study, the scout-guided model showed consistent performance across repeat scans.

CONCLUSIONS: By leveraging routinely acquired scout scans, this approach enables accurate quantitative PET reconstruction without a full CT, substantially reducing radiation dose and misalignment artifacts.