Analytical Performance of a Novel Latex Turbidimetric Immunoassay, "Nanopia TARC", for TARC/CCL17 Measurement: A Retrospective Observational Study.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Yamashita Keita, Takebayashi Shiori, Murata Wataru, Hirai Nao, Ito Yui, Mitsui Mayuka, Saito Mina, Sato Kei, Terada Miyuki, Niizeki Noriyasu, Suzuki Akira, Ogitani Kenya, Fujikawa Toshihiko, Komori Marie, Inoue Nozomi, Arai Norimitsu, Maekawa Masato

Аннотация

Thymus- and activation-regulated chemokine (TARC, also known as CCL17) is used as a biomarker for atopic dermatitis. The methods currently used for its measurement are complex, time-consuming, and require large machinery, warranting the need for a method that is simple, has a quick turnaround time, and requires less complex machinery. We evaluated the analytical performance of a novel latex turbidimetric immunoassay method, "Nanopia TARC", on 174 residual serum samples from patients with skin or allergic diseases. This evaluation included the assessment of the limit of blank/detection/quantification (LOB/D/Q), precision, accuracy, linearity, interference, and commutability between Nanopia TARC and "HISCL TARC", based on the chemiluminescent enzyme immunoassay (CLEIA) method. The LOB/D/Q values were 13, 57, and 141 pg/mL, respectively. The coefficient of variation of the repeatability was 0.9-3.8%, and that of the intermediate precision was 2.1-5.4%. The total error of the accuracy was 1.9-13.4%. The linearity was 141 and 19,804 pg/mL for TARC. The correlation coefficient between Nanopia TARC and HISCL TARC determined using the Passing-Bablok regression analysis was 0.999. Furthermore, the concordance of diagnostic criteria with AD was 92%. Nanopia TARC was confirmed to have the same analytical performance for TARC measurement as the existing CLEIA method.

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты