MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1161/circulationaha.109.849885

Медицинская библиотека · Circulation · 2009

Percutaneous mitral annuloplasty for functional mitral regurgitation: results of the CARILLON Mitral Annuloplasty Device European Union Study.

Schofer Joachim, Siminiak Tomasz, Haude Michael, Herrman Jean P, Vainer Jindra, Wu Justina C, Levy Wayne C, Mauri Laura, Feldman Ted, Kwong Raymond Y, Kaye David M, Duffy Stephen J, Tübler Thilo, Degen Hubertus, Brandt Mathias C, Van Bibber Rich, Goldberg Steve, Reuter David G, Hoppe Uta C

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Functional mitral regurgitation (FMR), a well-recognized component of left ventricular remodeling, is associated with increased morbidity and mortality in heart failure patients. Percutaneous mitral annuloplasty has the potential to serve as a therapeutic adjunct to standard medical care.

METHODS AND

RESULTS

Patients with dilated cardiomyopathy, moderate to severe FMR, an ejection fraction <40%, and a 6-minute walk distance between 150 and 450 m were enrolled in the CARILLON Mitral Annuloplasty Device European Union Study (AMADEUS). Percutaneous mitral annuloplasty was achieved through the coronary sinus with the CARILLON Mitral Contour System. Echocardiographic FMR grade, exercise tolerance, New York Heart Association class, and quality of life were assessed at baseline and 1 and 6 months. Of the 48 patients enrolled in the trial, 30 received the CARILLON device. Eighteen patients did not receive a device because of access issues, insufficient acute FMR reduction, or coronary artery compromise. The major adverse event rate was 13% at 30 days. At 6 months, the degree of FMR reduction among 5 different quantitative echocardiographic measures ranged from 22% to 32%. Six-minute walk distance improved from 307+/-87 m at baseline to 403+/-137 m at 6 months (P<0.001). Quality of life, measured by the Kansas City Cardiomyopathy Questionnaire, improved from 47+/-16 points at baseline to 69+/-15 points at 6 months (P<0.001).

CONCLUSIONS

Percutaneous reduction in FMR with a novel coronary sinus-based mitral annuloplasty device is feasible in patients with heart failure, is associated with a low rate of major adverse events, and is associated with improvement in quality of life and exercise tolerance.

Для цитирования: Schofer Joachim, Siminiak Tomasz, Haude Michael, Herrman Jean P, Vainer Jindra, Wu Justina C, Levy Wayne C, Mauri Laura, Feldman Ted, Kwong Raymond Y, Kaye David M, Duffy Stephen J, Tübler Thilo, Degen Hubertus, Brandt Mathias C, Van Bibber Rich, Goldberg Steve, Reuter David G, Hoppe Uta C. Percutaneous mitral annuloplasty for functional mitral regurgitation: results of the CARILLON Mitral Annuloplasty Device European Union Study.. Circulation. 2009. DOI: 10.1161/circulationaha.109.849885.