Аннотация

PURPOSE: Minimally invasive mitral valve surgery (MIMVS) and transcatheter edge-to-edge repair (TEER) are complex procedures used to treat mitral valve (MV) pathologies, but with limited training opportunities available. To enable training, a realistic hemodynamic environment is needed. In this work we aimed to develop and validate a simulator that enables investigation of MV pathologies and their repair by MIMVS and TEER in a hemodynamic setting.

METHODS

Different MVs were installed in the simulator, and pressure, flow, and transesophageal echocardiographic measurements were obtained. To confirm the simulator's physiological range, we first installed a biological prosthetic, a mechanical prosthetic, and a competent excised porcine MV. Subsequently, we inserted two porcine MVs-one with induced chordae tendineae rupture and the other with a dilated annulus, along with a patient-specific silicone valve extracted from echocardiography with bi-leaflet prolapse. Finally, TEER and MIMVS procedures were conducted by experts to repair the MVs.

RESULTS

Systolic pressures, cardiac outputs, and regurgitations volumes (RVol) with competent MVs were 119 ± 1 mmHg, 4.78 ± 0.16 l min-1, and 5 ± 3 ml respectively, and thus within the physiological range. In contrast, the pathological MVs displayed increased RVols. MIMVS and TEER resulted in a decrease in RVols and mitigated the severity of mitral regurgitation.

CONCLUSION

Ex-vivo modelling of MV pathologies and repair procedures using the described simulator realistically replicated physiological in-vivo conditions. Furthermore, we showed the feasibility of performing MIMVS and TEER at the simulator, also at patient-specific level, thus providing new clinical perspectives in terms of training modalities and personalized planning.

Для цитирования: Karl Roger, Romano Gabriele, Marx Josephin, Eden Matthias, Schlegel Philipp, Stroh Lubov, Fischer Samantha, Hehl Maximilian, Kühle Reinald, Mohl Lukas, Karck Matthias, Frey Norbert, De Simone Raffaele, Engelhardt Sandy. An ex-vivo and in-vitro dynamic simulator for surgical and transcatheter mitral valve interventions.. International journal of computer assisted radiology and surgery. 2023. DOI: 10.1007/s11548-023-03036-4.