Measurement of dose enhancement factor for Xoft Axxent electronic brachytherapy device using nanoparticle-embedded alginate film and radiochromic film.

Kakade Nitin R, Das Abhishek, Kumar Rajesh, Sharma S D, Maiti Nandita, Chadha Ridhima, Sapra B K

СтатьяПеревод пока не подготовлен

Аннотация

AIM: Inw nanoparticles-aided radiotherapy, the radiation sensitivity of tumor is increased with the infusion of nanoparticles in tumor. This therapeutic modality is capable of delivering enhanced dose to tumor, without exceeding the normal tissue tolerance dose. Further, the quantification of the enhanced dose using suitable dosimeter is important. The present study is aimed at measuring the dose enhancement factors (DEFs) using the combination of nanoparticles-embedded alginate (Alg) film and unlaminated Gafchromic EBT3 film.

MATERIALS AND

METHODS

Gold nanoparticles (AuNPs)- and silver nanoparticles (AgNPs)-embedded Alg polymer films were synthesized and characterized using standard techniques. Further, a customized version of the Gafchromic EBT3 film, i.e., unlaminated EBT3 film, was specially fabricated. The DEFs were measured using Xoft Axxent electronic brachytherapy device.

RESULTS

The surface plasmon resonance (SPR) and particle size of AuNPs were found to be 550 and 15 ± 2 nm, respectively. In the case of AgNPs, the SPR and particle size were recorded as 400 and 13 ± 2 nm, respectively. The DEFs measured, using unlaminated EBT3 film, for Xoft Axxent electronic brachytherapy using AuNPs and AgNPs were 1.35 ± 0.02 and 1.20 ± 0.01, respectively.

CONCLUSION

The increase in dose enhancement during nanoparticles-aided electronic brachytherapy can be attributed to dominance of photoelectric effect, due to the presence of low-energy X-rays. The investigation indicates that the Xoft Axxent electronic brachytherapy device is suitable for nanoparticles-aided brachytherapy.

Для цитирования: Kakade Nitin R, Das Abhishek, Kumar Rajesh, Sharma S D, Maiti Nandita, Chadha Ridhima, Sapra B K. Measurement of dose enhancement factor for Xoft Axxent electronic brachytherapy device using nanoparticle-embedded alginate film and radiochromic film.. Journal of cancer research and therapeutics. 2023. DOI: 10.4103/jcrt.jcrt_207_21.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.