Conjunctival bleb tearing by XEN gel stent after conjunctival compression sutures.

Montolío-Marzo Santiago, Lanzagorta-Aresti Aitor, Piá-Ludeña Jose Vicente, Davó-Cabrera Juan María

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: As minimally invasive glaucoma surgery devices emerge, newer complications are reported. XEN gel stent is safer than classic glaucoma surgery but may also lead to severe adverse effects as hypotony. Compression conjunctival sutures are a useful treatment for overfiltration hypotony after XEN gel stent implantation, but might force bleb tearing by the implant as it gets tightly close to the sutured conjunctiva. This complication has not been previously reported.

METHODS

We report a patient with overfiltration hypotony after XEN gel implantation. Conjunctival compression sutures and implant relocation ab interno were performed obtaining a good outcome. As conjunctiva got tightly close to the stent, the device eroded the bleb so leakage and implant extrusion were found.

RESULTS

Bleb reconstruction by conjunctival autograft was performed in order to avoid hypotony due to continuous conjunctival leakage.

CONCLUSION

Conjunctival compression sutures may be used for overfiltration control after XEN gel implantation but we must be aware of potential conjunctival damage due to implant erosion even if it is correctly positioned. Close follow-up and early suture release may be convenient in these patients.

Для цитирования: Montolío-Marzo Santiago, Lanzagorta-Aresti Aitor, Piá-Ludeña Jose Vicente, Davó-Cabrera Juan María. Conjunctival bleb tearing by XEN gel stent after conjunctival compression sutures.. European journal of ophthalmology. 2020. DOI: 10.1177/1120672120970862.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.