DOI 10.1016/j.jcv.2017.08.015

Медицинская библиотека · Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2017

A European multicientre study on the comparison of HBV viral loads between VERIS HBV assay and Roche COBAS® TAQMAN® HBV test, Abbott RealTime HBV assay, Siemens VERSANT HBV assay, and Qiagen artus HBV RG kit.

Braun Patrick, Delgado Rafael, Drago Monica, Fanti Diana, Fleury Hervé, Izopet Jacques, Lombardi Alessandra, Marcos MaAngeles, Sauné Karine, O'Shea Siobhan, Pérez-Rivilla Alfredo, Ramble John, Trimoulet Pascale, Vila Jordi, Whittaker Duncan, Artus Alain, Rhodes Daniel

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Hepatitis B viral load testing is essential to treatment and monitoring decisions in patients with chronic Hepatitis B. Beckman Coulter has developed the VERIS HBV Assay (Veris) for use on the fully automated DxN VERIS Molecular Diagnostics System.1

OBJECTIVES

To evaluate the clinical performance of the Veris HBV Assay at multiple EU laboratories STUDY

DESIGN

Method comparison was performed with a total of 344 plasma specimens from HBV infected patients tested with Veris and COBAS® TaqMan® HBV Test (Cobas), 207 specimens tested with Veris and RealTime HBV Assay (RealTime), 86 specimens tested with Veris and VERSANT® HBV Assay (Versant), and 74 specimens tested with Veris and artus® HBV RG PCR kit (artus).

RESULTS

Bland-Altman analysis showed average bias of -0.46 log10 IU/mL between Veris and Cobas, -0.46 log10IU/mL between Veris and RealTime, -0.36 log10IU/mL between Veris and Versant, and -0.12 log10IU/mL between Veris and artus. Bias was consistent across the assay range. Patient monitoring results using Veris demonstrated similar viral load trends over time to Cobas, RealTime, and artus.

CONCLUSIONS

The VERIS HBV Assay demonstrated comparable clinical performance, with varying degrees of negative bias, compared to other currently marketed assays for HBV DNA monitoring. This negative bias should be taken into consideration if switching monitoring methods to Veris.

Для цитирования: Braun Patrick, Delgado Rafael, Drago Monica, Fanti Diana, Fleury Hervé, Izopet Jacques, Lombardi Alessandra, Marcos MaAngeles, Sauné Karine, O'Shea Siobhan, Pérez-Rivilla Alfredo, Ramble John, Trimoulet Pascale, Vila Jordi, Whittaker Duncan, Artus Alain, Rhodes Daniel. A European multicientre study on the comparison of HBV viral loads between VERIS HBV assay and Roche COBAS® TAQMAN® HBV test, Abbott RealTime HBV assay, Siemens VERSANT HBV assay, and Qiagen artus HBV RG kit.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2017. DOI: 10.1016/j.jcv.2017.08.015.