BACKGROUND

Accidental catheter removal or drain dislodgment, including tube thoracostomy, is a common, high-risk complication in hospitalized and ambulatory patients that often necessitates an additional procedure, increased length of stay, and increased cost.

METHODS

The aim of this study was to compare the tensile strength of pigtail catheter fixation using a simple interrupted suture, a U-stitch suture, or 2 simple interrupted skin sutures in a standardized skin model. Catheters were sutured to the skin, penetrating the collagen layer, with 1 of the 3 suture techniques and varying suture combinations.

RESULTS

For each trial, breakage occurred at the suture or knot. The mean breakpoint varied significantly between 2 simple interrupted sutures and both the 1 simple interrupted suture and the U-stitch technique (analysis of variance post hoc test P < .001), with the 2 simple interrupted suture technique withholding nearly 40% more force. Using the strongest suture, 0 silk, on a deceased adult sheep to secure a pigtail thoracostomy catheter yielded identical data compared with the standardized skin model.

CONCLUSIONS

In conclusion, 2 simple interrupted skin sutures to secure a pigtail catheter has very low risk with a strongly positive benefit.

Для цитирования: Mashni Peyton, Savage Clare, Jax Michael, Reiter Anna Maria, Butler Gordon, Zwischenberger Joseph B. The Catheter Fell Out.. Annals of thoracic surgery short reports. 2024. DOI: 10.1016/j.atssr.2024.07.009.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.