World journal of urology · 2022
The effect of prolonged laser activation on irrigation fluid temperature: an in vitro experimental study.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Tsaturyan Arman, Peteinaris Angelis, Pantazis Lampros, Farsari Ergina, Pagonis Konstantinos, Adamou Constantinos, Vagionis Athanasios, Natsos Anastasios, Liatsikos Evangelos, Kallidonis Panagiotis
Аннотация
PURPOSE: To investigate the effect of prolonged laser activation on irrigation fluid temperature by varying the power settings flow rate (10-30 ml/min).
MATERIALS AND METHODS: An experimental study using a 20 ml syringe, 12/14 ureteral access sheath, a dual-lumen catheter and a thermocouple was performed. The laser was fired with 12 W (0.3 J × 40 Hz), 40 W (1 J × 40 Hz), 60 W (1.5 J × 40 Hz) using Quanta Ho 150 W (Quanta System, Samarate, Italy). All trials were performed with fluid outflow rate of 10, 20 and 30 ml/min with the fixed fluid volume at 10 ml.
RESULTS: Continuous laser activation for 10 min with the outflow rate of 10 ml/min using only 12 W resulted to continuous temperature rise to as high as 83 °C. Similar rise of temperatures were observed for 40 W and 60 W with 10 ml/min outflow rate with intermittent laser activation. With 20 and 30 ml/min outflow rates the maximum temperatures for all power settings were below the threshold (< 43 °C). However, the time to reach the same total emitted energy was 60% and 40% shorter 60 W and 40 W, respectively.
CONCLUSION: Our study found that continuous laser activation with as less as 12 W using 10 ml/min outflow rate increased the irrigation fluid temperature above the threshold only after 1 min. In the current experimental setup, with the fluid outflow rate of 20 and 30 ml/min safe laser activation with 60 W and 40 W (temperature < 43 °C) can be achieved reaching the same total emitted energy as with 12 W in significantly shorter time period.
MATERIALS AND METHODS: An experimental study using a 20 ml syringe, 12/14 ureteral access sheath, a dual-lumen catheter and a thermocouple was performed. The laser was fired with 12 W (0.3 J × 40 Hz), 40 W (1 J × 40 Hz), 60 W (1.5 J × 40 Hz) using Quanta Ho 150 W (Quanta System, Samarate, Italy). All trials were performed with fluid outflow rate of 10, 20 and 30 ml/min with the fixed fluid volume at 10 ml.
RESULTS: Continuous laser activation for 10 min with the outflow rate of 10 ml/min using only 12 W resulted to continuous temperature rise to as high as 83 °C. Similar rise of temperatures were observed for 40 W and 60 W with 10 ml/min outflow rate with intermittent laser activation. With 20 and 30 ml/min outflow rates the maximum temperatures for all power settings were below the threshold (< 43 °C). However, the time to reach the same total emitted energy was 60% and 40% shorter 60 W and 40 W, respectively.
CONCLUSION: Our study found that continuous laser activation with as less as 12 W using 10 ml/min outflow rate increased the irrigation fluid temperature above the threshold only after 1 min. In the current experimental setup, with the fluid outflow rate of 20 and 30 ml/min safe laser activation with 60 W and 40 W (temperature < 43 °C) can be achieved reaching the same total emitted energy as with 12 W in significantly shorter time period.