Safety and efficacy of hybrid closed-loop insulin pump therapy in adolescents and adults with type 1 diabetes in China.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Shen Jing, Zeng Tianshu, Li Xia, Zheng Chao, Li Yufeng, Wu Xiaohong, Yao Hui, Chen Linqi, Liu Yang, Zhu Weiwei, Fu Junfen, Zhao Weigang, Mu Yiming
Аннотация
AIMS: To evaluate the effectiveness and safety of a hybrid closed-loop (HCL) insulin infusion system in treating adult and adolescent patients with type 1 diabetes (T1D).
MATERIALS AND METHODS: A total of 126 participants (22 adolescents and 104 adults) underwent a 2-week initial treatment phase followed by a 2-week run-in phase with the Equil® S HCL insulin pump system MTM-Z in Manual Mode with calibration-free continuous glucose monitoring (CGM) System G7, followed by a 12-week study phase with HCL Auto Mode enabled. The primary outcome was the overall mean change in CGM time in range (TIR; 70-180 mg/dL). Safety endpoints included rates of severe hypoglycaemia, hyperglycaemia and diabetic ketoacidosis (DKA).
RESULTS: Compared with the baseline run-in, Auto Mode increased TIR from 69.11% to 76.72% (p < 0.001) and reduced time below range (<70 mg/dL) from 3.92% to 2.73% (p < 0.001). The mean sensor glucose (SG) decreased from 151 ± 28 to 144 ± 19 mg/dL. The proportion of achieving TIR of >70% increased from 56.03% to 77.59%. The standard deviation (SD: -5.90 mg/dL) and coefficient of variation (CV: -2.49%) reduced significantly. The greatest change in TIR (mean 20.55%), mean SG (estimated treatment difference [ETD]: -30.13, 95% confidence interval [CI]: -41.84, -18.41; p = 0.001) and SD of SG (ETD: -14.21, 95% CI: -21.17, -6.71; p = 0.004) was observed in participants with a baseline glycosylated haemoglobin A1c (HbA1c) value of >8.0%. The percentage of achieving combined recommendations for time at SG ranges increased from 25.86% to 56.90%. No severe hypoglycaemia, DKA or adverse events were observed.
CONCLUSIONS: HCL therapy is safe in Chinese adolescent and adult patients with T1D. The 12-week application of the Equil® S HCL insulin pump system MTM-Z with Auto Mode increased TIR and reduced HbA1c, hyperglycaemia and hypoglycaemia.
MATERIALS AND METHODS: A total of 126 participants (22 adolescents and 104 adults) underwent a 2-week initial treatment phase followed by a 2-week run-in phase with the Equil® S HCL insulin pump system MTM-Z in Manual Mode with calibration-free continuous glucose monitoring (CGM) System G7, followed by a 12-week study phase with HCL Auto Mode enabled. The primary outcome was the overall mean change in CGM time in range (TIR; 70-180 mg/dL). Safety endpoints included rates of severe hypoglycaemia, hyperglycaemia and diabetic ketoacidosis (DKA).
RESULTS: Compared with the baseline run-in, Auto Mode increased TIR from 69.11% to 76.72% (p < 0.001) and reduced time below range (<70 mg/dL) from 3.92% to 2.73% (p < 0.001). The mean sensor glucose (SG) decreased from 151 ± 28 to 144 ± 19 mg/dL. The proportion of achieving TIR of >70% increased from 56.03% to 77.59%. The standard deviation (SD: -5.90 mg/dL) and coefficient of variation (CV: -2.49%) reduced significantly. The greatest change in TIR (mean 20.55%), mean SG (estimated treatment difference [ETD]: -30.13, 95% confidence interval [CI]: -41.84, -18.41; p = 0.001) and SD of SG (ETD: -14.21, 95% CI: -21.17, -6.71; p = 0.004) was observed in participants with a baseline glycosylated haemoglobin A1c (HbA1c) value of >8.0%. The percentage of achieving combined recommendations for time at SG ranges increased from 25.86% to 56.90%. No severe hypoglycaemia, DKA or adverse events were observed.
CONCLUSIONS: HCL therapy is safe in Chinese adolescent and adult patients with T1D. The 12-week application of the Equil® S HCL insulin pump system MTM-Z with Auto Mode increased TIR and reduced HbA1c, hyperglycaemia and hypoglycaemia.